T-Cube NanoTrak Autoalignment Controller
6.2
NanoTrak Settings Window
When the 'Settings' button on th e GUI pan el is clicked, the 'Settings' window is
displayed. This panel allows data such scan circle frequency and diameter, and loop
gain to be entered. The various parameters are described below.
6.2.1
Tracking tab
Fig. 6.2 NanoTrak Settings panel - Tracking Tab
Circle
Frequency
- the frequency of the scanning circle. The circle scanning frequency
lies in the range 17.5 to 87.5Hz. The factory default setting for the scanning
frequency is 43.75Hz. This means that a stage driven by the NanoTrak makes
43.75 circular movements per second. Different frequency settings allow more
than one NanoTrak to be used in the same alignment scenario. Refer to Section
5.3.6. for more information.
Diameter Mode
- this setting allows the different modes of circle diameter
adjustment to be enabled and disabled.
If the Diameter Mode is set to '
User
', the circle diameter remains at the value set
by the user in the
User Diameter
parameter. The diameter of the scanning circle
is set in NanoTrak units. The circle diameter is measured on a scale such that 10
units is the width and height of the screen. If, for example, the NanoTrak is driving
a 20 micron actuator, a circle diameter of 1 unit will result in a real diameter of 2
microns. The scanning circle diameter can have values in the range 0 to 5.0 with