![THORLABS T-Cube NanoTrak TNA001/IR Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/thorlabs/t-cube-nanotrak-tna001-ir/t-cube-nanotrak-tna001-ir_user-manual_1110784014.webp)
14
HA0165T Rev 4K Sept 2017
Chapter 3
3.3.2
Supply Voltage and Current Requirements
3.3.3
Connecting To A Standalone Power Supply
Fig. 3.3 Front Panel Power Supply Connector
1) Using the front panel connector as shown above, connect the unit to a regulated
DC power supply of the rating detailed in Section 3.3.2.
2) Switch on the Power Supply unit.
3) The Version number of the embedded software is displayed for a few seconds.
Supply
Minimum
Maximum
Max Operating Current
+5V
+4.9V
+5.1V
400 mA
+15V
+14.5V
+15.5V
250 mA
-15V
-14.5V
-15.5V
100 mA
Warning
The unit must be connected only to a DC supply as detailed in Section
3.3.2. Connection to a supply of a different rating may cause damage to
the unit and could result in injury to the operator.
Caution
Do not connect the T-Cube unit to a 'live' external power supply. Doing so (i.e.
“hot plugging”) carries the risk of PERMANENT damage to the unit. Always
ensure the power supply unit is isolated from the mains before connecting to
the T-Cube unit. Always power up the units by connecting their power supply
to the mains. Similarly, to power down the unit, disconnect the power supply
from the mains before disconnecting the T-Cube.
USB
POWER
DC +15V, -15V, +5V