28
HA0165T Rev 4K Sept 2017
Chapter 5
Analogue Input Input Mode -
Select
Analogue Channel 1
Loop Mode -
Select
Open Loop
14)If desired, click the ‘Persist Settings To Hardware’ box. These settings will then be
loaded on each power up cycle.
Fig. 5.3 Piezo Driver Settings
15)Click ‘OK to save the settings.
16)On the GUI panel, turn the ‘Output’ control to give a reading of 0V on the display.
17)Click the ‘Settings’ button on GUI of the Piezo Driver fitted in bay 2. The Settings
panel is displayed
Make the following parameter settings, as shown in Fig. 5.3
Output Voltage Range -
Set the corresponding voltage for the piezo being
driven
, 75V, 100V or 150V.
Input Settings -
Select
So External Signal
Analogue Input Input Mode -
Select
Analogue Channel 2
Loop Mode -
Select
Open Loop
18)If desired, click the ‘Persist Settings To Hardware’ box. These settings will then be
loaded on each power up cycle