THORLABS S4FC Series Скачать руководство пользователя страница 15

S4FC Series Fiber Coupled Laser Source

 

Chapter 5: Operation

 

October 24, 2017, Rev R 

Page 13

 

5.10.2. Command Line Operation 

The system may also be controlled by a command line language through the USB port. This is offered to enable 
operation through a terminal interface or for those who may want to write their own program to control the unit. 
The command language is described below. Prior to running the command line interface, the unit should be 
powered on, and a USB cable should be connected between the S4FC and the host. 

The terminal emulator should be set as follows: 

Terminal Emulator Settings 

Baud Rate 

115.2 kb/s

1

 

Data Bits 

Parity 

None 

Stop Bits 

Flow Control 

None 

 

If you are using a terminal program, be sure to check the option “Append line feeds to incoming line ends”, or a 
similar option, depending upon the program you are using. In HyperTerminal you can find this option by 
navigating to File 

 Properties, selecting the Settings Tab and clicking “ASCII Setup…” button. It is the first 

option in the “ASCII Receiving” box. 

If the connection is correct you will see the following after pressing the Enter key. 

Command error CMD_NOT_DEFINED  

Followed immediately by the prompt: > 

The basic structure of the interface is a keyword followed by either an equals sign “=” or a question mark “?”. The 
“=” or “?” will determine if the string is a command or a query. All strings (commands and queries) must be 
terminated by a carriage return (CR) or pressing the ENTER key on the computer. 

The command structure is as follows: 

Keyword = argument (CR) 

Where “keyword” defines the function and “argument” is a numerical value followed by a carriage return (CR). 

See listing below. 

The query structure is as follows: 

Keyword? (CR) 

The “keyword” defines the function and the question mark (?) indicates a query. The string is terminated with a 
carriage return (CR). See listing below. There are a few exceptions to this which are noted below. 

The prompt symbol “>” will appear on power up and after a command is accepted by the system indicating it is 
ready to receive another command line. 

 

 

                                                      

1

 kilobits per second, other baud rates can be applied. 

Содержание S4FC Series

Страница 1: ...S4FC Series Fiber Coupled Laser Sources User Guide...

Страница 2: ...utput Power 8 5 7 Enabling the Laser 9 5 8 Turning the Laser Off 10 5 9 Modulating the Laser Output 10 5 10 Computer Controlled Operation 11 5 10 1 GUI Operation 11 5 10 2 Command Line Operation 13 5...

Страница 3: ...e Symbol Description Direct Current Alternating Current Both Direct and Alternating Current Earth Ground Terminal Protective Conductor Terminal Frame or Chassis Terminal Equipotentiality On Supply Off...

Страница 4: ...C wall socket Do not replace detachable mains lead with an inferior cord This may cause a failure damage shock or injury Changing of the mains fuse can be done by the user refer to section 7 3 on page...

Страница 5: ...iber care Please see section 5 2 page 7 regarding cleaning of fiber tips in order to prevent fiber damage and system degradation LASER RADIATION AVOID EXPOSURE TO BEAM CLASS 3B LASER RODUCT 660 nm 65...

Страница 6: ...ased on an internal photodiode monitor and temperature are set by use of rotary adjustment knobs located on the front panel Tactile buttons allow for the selection of either laser power or temperature...

Страница 7: ...ts are not obstructed 2 Make sure the POWER key switch on the front of the unit is in the OFF position key perpendicular to working surface 3 Plug the female end of the AC line cord provided into the...

Страница 8: ...nector Laser Output Cover Cover should be replaced when not in use Laser Emission Indicator Blinks for 5 seconds before laser turns ON Power Adjust Knob 4 Digit Display Laser Enable Button Press to ac...

Страница 9: ...ey on the fiber patch cable connector with the alignment slot on the S4FC output aperture Gently insert the fiber and thread the outer sleeve to the output aperture threads until finger tight Make sur...

Страница 10: ...ED labeled C next to the 4 digit LED display will be green when the temperature is within 1 C The C LED will turn amber whenever the temperature moves more than 1 C from the set point even if it is in...

Страница 11: ...to lase This start up value can be changed by a terminal command or from the GUI If the laser s Start From Last value is set to 1 which saves the last power setting then this option is superseded in...

Страница 12: ...in mind when using the modulation input since it will limit the actual input voltage range The second mode From Zero allows the analog modulation to utilize the full 0 5 volt input range This allows...

Страница 13: ...same directory on your storage device Figure 4 File locations of GUI software The necessary drivers can also be found and installed from the provided CD After inserting the CD the window in Figure 5 w...

Страница 14: ...mit the output by setting lower current than the default max current value Enable Enables the LD and sets the output power See section 5 7 page 9 Set Point Indicates or sets the temperature set point...

Страница 15: ...ab and clicking ASCII Setup button It is the first option in the ASCII Receiving box If the connection is correct you will see the following after pressing the Enter key Command error CMD_NOT_DEFINED...

Страница 16: ...C HEATSINK TEMP Returns TEC heat sink temp in C AMBIENT TEMP Returns ambient temp in C INTERLOCK Returns interlock 0 Inactive 1 Active ENABLE Returns LD enabled state 0 Off 1 On ENABLE Sets LD enabled...

Страница 17: ...a shorting device installed in the Interlock connector If you are not going to use this feature then you can leave the shorting device installed and the unit will operate normally as described in the...

Страница 18: ...are over the fiber core WARNING DO NOT connect a fiber patch cord to the laser s FC PC bulkhead connector without cleaning the fiber tips first DO NOT connect a fiber patch cord to the unit while the...

Страница 19: ...w the AC power cord connection on the rear panel of the unit Carefully use a flat blade screwdriver to open the fuse tray Remove the existing fuse and install the appropriate fuse Specification Value...

Страница 20: ...le E016 Interlock Tripped The interlock circuit is open If an interlock setup is being used resolve the tripped interlock Otherwise make sure the interlock shorting pin is in place Then re enable the...

Страница 21: ...Slot Use a Viewing Card to safely determine laser output Unit is enabled but there is no Output Analog Modulation 1 Check the output of the modulating source and confirm the Output is On 2 If using A...

Страница 22: ...20 nm SMF 28 Ultra 3B 48 50 0 004 0 1 0 009 0 2 General Specifications Display Accuracy 5 Laser Adjustment Resolution 0 01 mW While Laser is Enabled Laser Adjustment Range Default Adjusts from Laser T...

Страница 23: ...S4FC Series Fiber Coupled Laser Source Chapter 10 Mechanical Drawing October 24 2017 Rev R Page 21 Chapter 10 Mechanical Drawing 3 00 76 1 mm 12 47 316 6 mm 5 92 150 3 mm...

Страница 24: ...other Thorlabs products such as Pure OEM products that means assemblies to be built into a unit by the user e g OEM laser driver cards Components Mechanics and optics Left over parts of units disassem...

Страница 25: ...S4FC Series Fiber Coupled Laser Source Chapter 12 Certificate of Compliance October 24 2017 Rev R Page 23 Chapter 12 Certificate of Compliance...

Страница 26: ...3 6 3 Kitamachi Nerima ku Tokyo 179 0081 Japan Tel 81 3 6915 7701 Fax 81 3 6915 7716 www thorlabs co jp Email sales thorlabs jp UK and Ireland Thorlabs Ltd 1 Saint Thomas Place Ely CB7 4EX Great Brita...

Страница 27: ...www thorlabs com...

Отзывы: