S4FC Series Fiber Coupled Laser Source

 

Chapter 5: Operation

 

Page 10

 

TTN051387-D02 

 

1 = Threshold:

 Sets the starting value to the laser diodes threshold value. 

 

0 = Zero Power:

 Sets the starting value to nearly 0.1 mW power; or threshold if threshold is less than 

0.1 mW. 

5.8.  Turning the Laser Off 

WARNING 

DO NOT turn the power key switch off until laser is disabled. Failing to disable the laser first can 

damage the unit permanently. 

When completely powering down an enabled unit, the operator needs to disable the output first to prevent 
damage to the unit. Anytime the unit is turned OFF and then turned back ON, the laser will be disabled until the 
ENABLE switch is pressed. 

  Press the ENABLE button to turn OFF the laser. 

  (Optional) Turn the power key counterclockwise to the off potion to power down the unit. 

5.9.  Modulating the Laser Output 

The MOD IN input can be used to modulate the laser output or set the laser output remotely using a 5 V power 
source. The 5 V maximum input corresponds to the maximum power, which operates using a constant current 
drive technique. The resulting actual output power is dependent on the set current and operating temperature. In 
addition, in order to eliminate a dead zone in the Power Adjustment knob (due to the lasing threshold current), the 
output of the unit is offset to the threshold current of the coupled laser diode. Adjusting the knob below threshold 
will immediately set the current to almost 0 mA. 
 
Therefore there are two modes of modulation available in the control software. In the default mode, 

From

 

Threshold

, the control knob adjusts the laser from threshold to Max power. The analog Modulation voltage will be 

limited to less than 5 V, but a DC offset will not be required. This should be kept in mind when using the 
modulation input since it will limit the actual input voltage range. The second mode, 

From Zero

, allows the analog 

modulation to utilize the full 0 – 5 volt input range. This allows more flexibility with the drawback that a minimum 
voltage or DC offset will be required to operate above the laser threshold. 

 

  Connect a signal generator or 0 to 5 V power source to the unit using a BNC type connector. 

  Set the PWR ADJ knob on the front panel to its full counter clockwise setting, or increase slightly to the 

threshold current mode. 

  Press the ENABLE switch to turn on the laser, wait for the safety delay to time out. 

For Threshold Offset 

- To calibrate your input apply a DC voltage to the MOD IN and slowly increase until the 

display reaches the maximum rated power, if exceeded the system will limit and may disable. This will be the max 
voltage allowable. Then apply the appropriate signal to the MOD IN input. Increase to 5V, if too much voltage is 
applied internal current limiting will prevent damage to the laser diode and the laser diodes will be immediately 
disabled. If an internal threshold is set the acceptable drive voltage range will be less than 5V. 

 

For Standby Operation 

– Apply and appropriate signal between 0 and 5 V to the MOD IN for the desired 

channel. Determine the DC offset by increasing the voltage slowly until a large power jump is seen on the output. 
An easier method it to apply a 1 Vpp sinusoid and adjust the DC offset until the sinusoid appears and then the 
bottom edge is no longer clipping. This method requires applying the laser output to a photo-detector and 
monitoring with an oscilloscope. Next maximize the amplitude so that the peak is 5 V or less as desired. 
 

Modulation Maximum Frequency 

– The maximum modulation frequency is 100 kHz. Normally a sine wave is 

used for modulation, if a square wave modulation is used instead; consider the bandwidth as a function of the rise 
time:  

 

0.35

 

 

Содержание S4FC Series

Страница 1: ...S4FC Series Fiber Coupled Laser Sources User Guide...

Страница 2: ...utput Power 8 5 7 Enabling the Laser 9 5 8 Turning the Laser Off 10 5 9 Modulating the Laser Output 10 5 10 Computer Controlled Operation 11 5 10 1 GUI Operation 11 5 10 2 Command Line Operation 13 5...

Страница 3: ...e Symbol Description Direct Current Alternating Current Both Direct and Alternating Current Earth Ground Terminal Protective Conductor Terminal Frame or Chassis Terminal Equipotentiality On Supply Off...

Страница 4: ...C wall socket Do not replace detachable mains lead with an inferior cord This may cause a failure damage shock or injury Changing of the mains fuse can be done by the user refer to section 7 3 on page...

Страница 5: ...iber care Please see section 5 2 page 7 regarding cleaning of fiber tips in order to prevent fiber damage and system degradation LASER RADIATION AVOID EXPOSURE TO BEAM CLASS 3B LASER RODUCT 660 nm 65...

Страница 6: ...ased on an internal photodiode monitor and temperature are set by use of rotary adjustment knobs located on the front panel Tactile buttons allow for the selection of either laser power or temperature...

Страница 7: ...ts are not obstructed 2 Make sure the POWER key switch on the front of the unit is in the OFF position key perpendicular to working surface 3 Plug the female end of the AC line cord provided into the...

Страница 8: ...nector Laser Output Cover Cover should be replaced when not in use Laser Emission Indicator Blinks for 5 seconds before laser turns ON Power Adjust Knob 4 Digit Display Laser Enable Button Press to ac...

Страница 9: ...ey on the fiber patch cable connector with the alignment slot on the S4FC output aperture Gently insert the fiber and thread the outer sleeve to the output aperture threads until finger tight Make sur...

Страница 10: ...ED labeled C next to the 4 digit LED display will be green when the temperature is within 1 C The C LED will turn amber whenever the temperature moves more than 1 C from the set point even if it is in...

Страница 11: ...to lase This start up value can be changed by a terminal command or from the GUI If the laser s Start From Last value is set to 1 which saves the last power setting then this option is superseded in...

Страница 12: ...in mind when using the modulation input since it will limit the actual input voltage range The second mode From Zero allows the analog modulation to utilize the full 0 5 volt input range This allows...

Страница 13: ...same directory on your storage device Figure 4 File locations of GUI software The necessary drivers can also be found and installed from the provided CD After inserting the CD the window in Figure 5 w...

Страница 14: ...mit the output by setting lower current than the default max current value Enable Enables the LD and sets the output power See section 5 7 page 9 Set Point Indicates or sets the temperature set point...

Страница 15: ...ab and clicking ASCII Setup button It is the first option in the ASCII Receiving box If the connection is correct you will see the following after pressing the Enter key Command error CMD_NOT_DEFINED...

Страница 16: ...C HEATSINK TEMP Returns TEC heat sink temp in C AMBIENT TEMP Returns ambient temp in C INTERLOCK Returns interlock 0 Inactive 1 Active ENABLE Returns LD enabled state 0 Off 1 On ENABLE Sets LD enabled...

Страница 17: ...a shorting device installed in the Interlock connector If you are not going to use this feature then you can leave the shorting device installed and the unit will operate normally as described in the...

Страница 18: ...are over the fiber core WARNING DO NOT connect a fiber patch cord to the laser s FC PC bulkhead connector without cleaning the fiber tips first DO NOT connect a fiber patch cord to the unit while the...

Страница 19: ...w the AC power cord connection on the rear panel of the unit Carefully use a flat blade screwdriver to open the fuse tray Remove the existing fuse and install the appropriate fuse Specification Value...

Страница 20: ...le E016 Interlock Tripped The interlock circuit is open If an interlock setup is being used resolve the tripped interlock Otherwise make sure the interlock shorting pin is in place Then re enable the...

Страница 21: ...Slot Use a Viewing Card to safely determine laser output Unit is enabled but there is no Output Analog Modulation 1 Check the output of the modulating source and confirm the Output is On 2 If using A...

Страница 22: ...20 nm SMF 28 Ultra 3B 48 50 0 004 0 1 0 009 0 2 General Specifications Display Accuracy 5 Laser Adjustment Resolution 0 01 mW While Laser is Enabled Laser Adjustment Range Default Adjusts from Laser T...

Страница 23: ...S4FC Series Fiber Coupled Laser Source Chapter 10 Mechanical Drawing October 24 2017 Rev R Page 21 Chapter 10 Mechanical Drawing 3 00 76 1 mm 12 47 316 6 mm 5 92 150 3 mm...

Страница 24: ...other Thorlabs products such as Pure OEM products that means assemblies to be built into a unit by the user e g OEM laser driver cards Components Mechanics and optics Left over parts of units disassem...

Страница 25: ...S4FC Series Fiber Coupled Laser Source Chapter 12 Certificate of Compliance October 24 2017 Rev R Page 23 Chapter 12 Certificate of Compliance...

Страница 26: ...3 6 3 Kitamachi Nerima ku Tokyo 179 0081 Japan Tel 81 3 6915 7701 Fax 81 3 6915 7716 www thorlabs co jp Email sales thorlabs jp UK and Ireland Thorlabs Ltd 1 Saint Thomas Place Ely CB7 4EX Great Brita...

Страница 27: ...www thorlabs com...

Отзывы: