Chapter 3 Mechanical Installation
Chapter 3 Mechanical Installation
3.1 Unpacking
3.2 Environmental Conditions
Location
Indoor use only
Maximum altitude
2000 m
Temperature range
5
o
C to 40
o
C
Maximum Humidity
Less than 80% RH (non-condensing) at 31°C
To ensure reliable operation the unit should not be exposed to corrosive agents or
excessive moisture, heat or dust.
If the unit has been stored at a low temperature or in an environment of high humidity,
it must be allowed to reach ambient conditions before being powered up.
The unit must not be used in an explosive environment.
3.3 Mounting to a Work Surface
3.3.1 General
When mounting the lab jack close to other equipment, ensure that the travel of the
moving platform is not obstructed. If equipment mounted on the moving platform is
driven against a solid object, damage to the internal mechanism could occur. The
range of travel is 2.0” (50 mm) vertically.
Caution
Once removed from its packaging, the MLJ150 lab jack is easily damaged by
mishandling. The unit should only be handled by its base, not by the motor or
any attachments to the moving platform.
Note
Retain the packing in which the unit was shipped, for use in future transportation.
Warning
Operation outside the following environmental limits may adversely affect
operator safety.
Warning
The safety of any system incorporating this equipment is the responsibility of
the person performing the installation.