Appendix A Preventive Maintenance
Page 32
Appendix A Preventive Maintenance
A.1 Safety Testing
PAT testing in accordance with local regulations, should be performed on a regular
basis, (typically annually for an instrument in daily use).
A.2 Cleaning
The fascia may be cleaned with a soft cloth, lightly dampened with water or a mild
detergent.
A.3 Troubleshooting
If an unexpected obstruction to the stage motion is encountered, the control system
interface software disables the current to the associated drive channel, which may or
may not result in the channel being disabled. After the obstruction to be removed, the
associated channel of the controller should first be disabled to clear any fault codes,
then re-enabled.
In the event of a breakdown, or malfunction of the product please contact Thorlabs
Tech Support. Contact details are contained in Appendix E.
Warning
The equipment contains no user servicable parts. There is a risk of electrical
shock if the equipment is operated with the covers removed. Only personnel
authorized by Thorlabs Ltd and trained in the maintenance of this equipment
should remove its covers or attempt any repairs or adjustments. Maintenance
is limited to safety testing and cleaning as described in the following sections.
Caution
The instrument contains a power supply filter. Insulation testing of the power
supply connector should be performed using a DC voltage.
Warning
Disconnect the power supply before cleaning the unit.
Never allow water to get inside the case.
Do not saturate the unit.
Do not use any type of abrasive pad, scouring powder or solvent,
e.g. alcohol or benzene.