Rev B Apr 2018
Page 30
Chapter 7 Software Reference
7.3.5 Advanced Tab
Fig. 7.8 MLJ150 Stage - Advanced Settings
Power Settings
The MLJ150 is designed to vary the phase powers (current) in the motor coils
depending on the operating state of the motor - moving or stationary. Typically, when
a stepper motor is at rest it is advisable to reduce the phase (holding) currents so that
the motor does not overheat. When moving, these phase currents are boosted to
provide sufficient motor torque and minimise the possibility of stalling (missed steps)
The moving phase powers are set automatically by the unit and cannot be adjusted.
The Resting Power is entered as a percentage of full power.
Note
The default values applied by the software have been selected based on the type of
stage or actuator associated with the motor drive. Modify these values with caution
(particularly the rest power) as the risk of damage to the motor due to overheating is
significant.