3.6 ITC5xx specific commands
ITC500 / page 75
3.6.7 Reading the laser diode hardware limit (LIMCP)
Syntax:
":LIMCP:ACT?"
Response:
[:LIMCP:ACT <NR3>]
Description: Queries the laser diode limit current (adjusted via potentiometer).
(Refer to Chapter 1.8.1, "Setting the limit of the laser current I
LD LIM
" on page 16)
3.6.8 Laser diode current (ILD)
Programming:
Syntax:
":ILD:SET <NR3>"
Valid Range: Depends on the instrument type.
Default Value: 0 A
Assumption: The operation mode is switched to constant current.
Description: Sets the laser diode current. Unit: [A].
Reading the set current:
Syntax:
":ILD:SET?"
Response:
[:ILD:SET <NR3>]
Description: Queries the programmed laser diode current. Unit: [A].
Reading the minimum current:
Syntax:
":ILD:MIN?"
Response:
[:ILD:MIN <NR3>]
Description: Queries the minimum allowed laser diode current. Unit: [A].
Reading the maximum current:
Syntax:
":ILD:MAX?"
Response:
[:ILD:MAX <NR3>]
Description: Queries the maximum allowed laser diode current. Unit: [A].
Reading the actual current:
Syntax:
":ILD:ACT?"
Response:
[:ILD:ACT <NR3>]
Description: Queries the actual laser diode current. Unit: [A].