2.9 Bias voltage of the monitor diode (BIAS)
ITC500 / page 37
The button
PD POL
repeatedly toggles between four different settings:
CG
(monitor diode cathode grounded)
CG + BIAS
(monitor diode cathode grounded with bias voltage)
AG
(monitor diode anode grounded)
AG + BIAS
(monitor diode anode grounded with bias voltage)
CG
. . .
Toggling values follows always this order. If the “wake up“ default is
AG
, the setting
after pressing
PD POL
once will be
AG + BIAS
. If the “wake up“ default is
AG +
BIAS
, the setting after pressing
PD POL
once will be
CG
, etc...
Attention
Before switching on the bias voltage make sure that the monitor
diode polarity is set correctly.
If the monitor diode is forward biased with a voltage the resulting
current can damage or even destroy the monitor diode (= the laser
module).