2.15 Setup and function of temperature controllers
ITC500 / page 47
2.15.4.5 I-Share disconnection
For some cases, particularly for slow thermal loads, this procedure will not work
satisfactory.
Please follow the steps listed below to obtain a better adjustment.
Attention
This procedure should only be done by qualified service personnel!
Hazardous voltages exist within the device. As shipped from the
factory, all parts carrying line voltage are covered to reduce the
likelihood of contact. Use only isolated tools and be careful not to
touch any Mains..
•
Pull the line plug
•
Remove the cover of the ITC5xx
•
Set jumper JP8 on the mainboard
•
Adjust the P- and D-share as usual
•
Remove the jumper JP8 again
•
Adjust the I-share at last
•
Reinstall the cover