![THORLABS ITC502 Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/thorlabs/itc502/itc502_operation-manual_1110749018.webp)
1.5 Technical Data
ITC500 / page 10
Temperature control Output
Load resistance
>
10 k
Ω
Output voltage
0 ...
±
10 V
Transmission coefficient IC sensors
50 mV / °C ± 5%
Transmission coeff. thermistor (20 k
Ω
/ 200 k
Ω
range) 500 mV/ k
Ω
/ 50 mV/k
Ω
± 5%
Temperature window protection
Setting range T
WIN
0.5 °C ... 20 °C
Setting range R
WIN
(20 k
Ω
/ 200 k
Ω
range)
50
Ω
... 2 k
Ω
/ 500
Ω
... 20 k
Ω
Computer Interface
Setting resolution
16 Bit
Measurement resolution
12 ... 18 Bit
1)
1.5.2 Technical data ITC510
General Data
Line voltage
100 V / 115 V / 230 V (-10%, +15 %) (fixed)
Line frequency
50 ... 60 Hz
Max. power consumption
150 VA
Supply mains overvoltage
Overvoltage category II (Cat II)
Operating temperature
2
0 ... +40
°
C
Storage temperature
-40
°
C ... +70
°
C
Relative Humidity
Max. 80% up to 31°C, decreasing to 50% at 40 °C
Operation altitude
< 3000 m
Pollution Degree (indoor use only)
2
Warm up time for rated accuracy
≤
10 min
Weight
<
7 kg
Dimensions
220 x 120 x 377 mm
3
1)
in High Resolution mode, at reduced measurement speed
2)
non condensing