![THORLABS ITC502 Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/thorlabs/itc502/itc502_operation-manual_1110749021.webp)
1.5 Technical Data
ITC500 / page 13
Thermistor (2 k
Ω
/ 20 k
Ω
range)
Measurement current
100
µ
A / 10
µ
A
Control range
10
Ω
... 19.99 k
Ω
/ 100
Ω
... 199.9 k
Ω
Setting resolution (manual control)
1
Ω
/ 10
Ω
Setting resolution (remote control)
0.3
Ω
/ 3
Ω
Measurement resolution (manual control)
1
Ω
/ 10
Ω
Measurement resolution (remote control)
0.1
Ω
/ 1
Ω
Accuracy
±
5
Ω
/
±
50
Ω
Temperature stability (24 hours)
1)
≤
0.5
Ω
/
≤
5
Ω
Temperature control input
Input resistance
10 k
Ω
Control voltage
-10 ... +10 V
Transmission coefficient IC-sensors
2 °C/V
±
5%
Transmission coefficient thermistor (20 k
Ω
/ 200 k
Ω
range)
0.2 k
Ω
/V / 2 k
Ω
/V
±
5%
Temperature control Output
Load resistance
>
10 k
Ω
Output voltage
0 ...
±
10 V
Transmission coefficient thermistor
500 mV/ k
Ω
/ 50 mV/k
Ω
± 5%
Transmission coefficient IC sensors
50 mV / °C ± 5%
Temperature window protection
Setting range T
WIN
0.5 °C ... 20 °C
Setting range R
WIN
(20 k
Ω
/ 200 k
Ω
range)
50
Ω
... 2 k
Ω
/ 500
Ω
... 20 k
Ω
Computer Interface
Setting resolution
16 Bit
Measurement resolution
12 ... 18 Bit
2)
1)
Due to the nonlinear conversion from
Ω
to °C the stability in °C depends on the operating conditions
and the characteristics of the thermistor. E.g. for a typical thermistor at a set point of 10k
Ω
(25°C), a
0.5
Ω
stability translates into about 1mK temperature stability. At a set point of 5k
Ω
(38°C), the stability
is about 2mK.
2)
in High Resolution mode, at reduced measurement speed