© Adolf Thies GmbH & Co. KG · Hauptstraße 76 · 37083 Göttingen · Germany
021814/03/21
Phone +49 551 79001-0 · Fax +49 551 79001-65 · [email protected] · www.thiesclima.com
Page 60 of 86
Regis-
ter ad-
dress
Parameter name
Unit
Multipli-
cator
Explanation
Data type
36209
Inclination angle
θ
Angle be-
tween the Z-axis and the vertical
(34819)
1
°
10
Value / 10
(1 decimal place,
e.g. 240=24.0°)
S32
36211
Inclination angle
ρ
Angle between the X-axis and
the horizontal plane (34821)
1
°
10
Value / 10
(1 decimal place,
e.g. 240=24.0°)
S32
36213
Inclination Angle
Φ
Angle be-
tween the Y-axis and the hori-
zontal plane (34823)
1
°
10
Value / 10
(1 decimal place,
e.g. 240=24.0°)
S32
36215
Frequency vibration measure-
ment, x-axis (34825)
1
Hz
100
Value / 100
(2 decimal place,
e.g 2400=24.00Hz)
U32
36217
Acceleration, x-axis (34827)
1
mg
1
Value
(no decimal place,
e.g. 24=24mg)
U32
36219
Frequency vibration measure-
ment, y-axis (34829)
1
Hz
100
Value / 100
(2 decimal place,
e.g. 2400=24.00Hz)
U32
36221
Acceleration, y-axis (34831)
1
mg
1
Value
(no decimal place,
e.g. 24=24mg)
U32
36223
Frequency vibration measure-
ment, z-axis (34833)
1
Hz
100
Value
(no decimal place,
e.g. 24=24mg)
U32
36225
Acceleration, z-axis (34835)
1
mg
1
Value
(no decimal place,
e.g. 24=24mg)
U32
36227
Sensor status
(34811)
1
1
value
(no decimal place,
bit coded, depend-
ence on the sensor)
U32
Table 5 :
MODBUS Input Register
1
: The numbers in brackets designate the register addresses which represent the same s. For example,
the wind velocity is on address 30001 and on address 35001.
2
: Measuring values of the wind velocity (without compensation) acquired by means of standard characteristic curve.
3
: The measurement values of the wind speed detected by means of calibration characteristic.
4
: Average wind speed, based on the selected measured values (see command MS).
5
: The measured values of the wind speed (without compensation) recorded by means of the standard characteristic curve.
6.
Standard deviation of the wind speed, based on the selected measured values (see command MS).
7
: Minimum / maximum wind speed based on the selected measured values (see command MS).
8
: The air pressure compensated wind speed, based on the measured values recorded by means of the calibration curve.
9
: Info: It is the directly measured, uncorrected frequency. Independent of the command FO.
Note:
Due to the uninterrupted sequence of the measured val-
ues from address 35001, the MODBUS Master can read all