Bit number Function
Description
Bit 17…19
Source mean value calcu-
lation
The bits 17 to 19 show the wind velocity used for the cal-
culation of the mean values.
000: Wind velocity calculated by the standard characteritic
curve.
001: Wind velocity calculated by the calibration charact.
curve.
010: Wind velocity calculated by the calibration table and
compensated concerning air pressure.
Bit 20...31
-
-
Table 1 :
Status Word
11.2 Commands and Description
The following table lists the available commands and the corresponding password for read-
ing and writing:
Command
Initial
value
Factory
setting
MODBUS
Register
address
Description
Password
Reading
1
/
Writing
2
Command BR
96
40005
Baud rate
Without
User
Command BT
0
40019
Bus termination
Without
User
Command CI
0/1
40013
Command interpreter
Without
User
Command CK
137
-
Password for writing ac-
cess to calibration tables
Without
User
or CK
Command FB
1
40001
Quick start mode
Without
User
Command FO
4
40025
Frequency output
Without
User
Command HP
500
40041
Target value for the tem-
perature control unit
Without
User
Command HT
0
40017
Mode of the heater control
Without
User
Command ID
0/1
40003
ID number
Without
User
Command KX
0
-
Key /password
Without
User or
KX
Command KY
0
40009
Key/password
Without
Without
Command MI
10
40027
Averaging interval
Without
User
1
-
Source mean value calcula-
tion
Without
User
Command OH
-
40021
Read operating parameters
Without
Without
Command RC
-
Read entry from calibration
table
Without
Without
Command RS
-
-
Reset
Without
User
Command SF
0
40043
Frame format
Without
User
1
The device version 4.3352.x0.400 is delivered with 0 (THIES- Interpreter), the version 4.3352.x0.401
is delivered with 1 (MODBUS RTU).
2
The device version 4.3352.x0.400 is delivered with ID0, the version 4.3352.x0.401 is delivered with
ID1. If the sensor includes an additional label with the ID, the ID of the label is valid.
3
For reading the entries of the calibration table please see chapter 9.3.3 Calibration table in the
MODBUS-RTU Interpreter.