![Thermo 3307 Скачать руководство пользователя страница 41](http://html.mh-extra.com/html/thermo/3307/3307_operating-and-maintenance-manual_1104186041.webp)
Steri-Cult Incubator ____________________________________________________________________________Servicing
11-2
11.2 Replacing the Fuses
Refer to Wiring Diagram 3307-71, page 1 of 2, Fuse Holder
Detail 2.
The following procedure requires accessing the
electronics area. Qualified personnel only should
perform this procedure. Wearing a electrostatic
discharge wrist band or other grounding device is
recommended.
To access the only replaceable fuse in the incubator, follow
the steps below.
1. Turn off the incubator’s power switch and disconnect the
power cord.
2. Remove the four screws from the left side panel. Set the
panel aside.
3. On the left side of the opening is the electronics area.
Locate the fuse. See Table 11-1 below for replacement
fuse specifications. Replace the blown fuse with a new
one.
4. Install the panel and secure with the screws.
5. Return the unit to service. If the fuse blows after restor-
ing power to the incubator, contact the Technical
Services Department.
11.3 Replacing the CO
2
and RH Air Filters
The following procedure requires accessing the
electronics area. Qualified personnel only should
perform this procedure. Wearing a electrostatic
discharge wrist band or other grounding device is
recommended.
The CO
2
and RH air filters are recommended to be
replaced annually. Follow the steps below to replace the filters.
1. Turn off the incubator’s power switch and disconnect the
power cord.
2. Remove the four screws from the left side panel. Set the
panel aside.
3. On the left side of the opening is the electronics area.
Locate the CO
2
air filter (Figure 11-1). Disconnect care-
fully from the tubing and replace.
5. In the right side of the opening is the humidity system.
Locate the RH air filter in the upper right corner (Figure
11-2). Disconnect from the tubing and replace.
6. Install the panel and secure with the screws.
7. Return the unit to service.
Fuse Voltage
Manufacturers Part #
Amperage
Rating
Rupture
Speed
IEC
Letter
Code
115VAC ACC
BUSS GMC-1.0A
1.0 Amp
Time-Lag
T
250VAC ACC
BUSS GMC-740ma
750mAmp
Time-Lag
T
230VAC ACC
BUSS GMC-500ma
0.5 Amp
Time-Lag
T
Table 11-1, Fuse Replacement Chart
Figure 11-1
Figure 11-2
Содержание 3307
Страница 16: ...Steri Cult Incubator ________________________________________________________________________Calibration 3 2 ...
Страница 20: ...Steri Cult Incubator _______________________________________________________________________Configuration 4 4 ...
Страница 21: ...Steri Cult Incubator _______________________________________________________________________Configuration 4 5 ...
Страница 22: ...Steri Cult Incubator _______________________________________________________________________Configuration 4 6 ...
Страница 42: ......
Страница 43: ......
Страница 44: ......
Страница 45: ......
Страница 46: ......
Страница 47: ......
Страница 48: ......
Страница 49: ......
Страница 50: ......
Страница 51: ......
Страница 52: ......