k. Setting a High RH Alarm Limit (tracking alarm)
The high RH alarm limit is the deviation from the humidity
setpoint which causes a high RH alarm. The setpoint is vari-
able from 5 to 20% RH The factory setting is 10% RH. When
the alarm is activated, the humidity system is disabled.
1. Press the Mode key until the Configuration indicator
lights.
2. Press the right arrow until RH HI LMT XX is displayed
in the message center.
3. Press up/down arrow to change the RH high alarm limit.
4. Press Enter to save the RH high alarm limit.
5. Press the Mode key to return to Run mode or right/left
to go to next/previous parameter.
l. Enabling RH Alarms to Trip Contacts
The high and low RH alarms can be programmed to trip
the remote alarm contacts. A setting of ON activates this abili-
ty. A setting of OFF will not allow the RH alarm to trip the con-
tacts. The factory setting is ON.
1. Press the Mode key until the Configuration indicator
lights.
2. Press the right arrow until RH RELAY ON/OFF is dis-
played in the message center.
3. Press up/down arrow to toggle the setting ON or OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left
to go to next/previous parameter.
m. Selecting a Primary Tank w/ Gas Guard Option
On units equipped with the Gas Guard option, a primary
tank can be selected. The primary tank will be either Tank 1 or
2. The factory setting is Tank1.
1. Press the Mode key until the Configuration indicator
lights.
2. Press the right arrow until TANK SELECT X is dis-
played in the message center.
3. Press up/down arrow to toggle setting between 1 and 2 .
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left
to go to next/previous parameter.
Steri-Cult Incubator ______________________________________________________________________Configuration
4 - 3
n. Disabling the Gas Guard System
On units equipped with the Gas Guard option, the Gas
Guard system may be turned ON or OFF if it is not in use. The
factory setting is ON.
1. Press the Mode key until the Configuration indicator
lights.
2. Press the right arrow until GAS GUARD ON/OFF is
displayed in the message center.
3. Press up/down arrow to toggle the setting ON or OFF.
4. Press Enter to save the setting.
5. Press the Mode key to return to Run mode or right/left
to go to next/previous parameter.
o. Setting a RS485 Communications Address
(1535 compatible only)
On units that have the RS485 option, direct communication
with the Model 1535 alarm system can be established. Each
piece of equipment connected to the Model 1535 must have a
unique address. An address of 0-24 can be entered for the incu-
bator. A setting of 0 is an invalid address that the 1535 will
ignore. The factory setting for the RS485 address is 0.
1. Press the Mode key until the Configuration indicator
lights.
2. Press the right arrow until 485 ADDRESS XX is dis-
played in the message center.
3. Press up/down arrow to move the RS485 address.
4. Press Enter to save the RS485 address.
5. Press the Mode key to return to Run mode or right/left
to go to next/previous parameter.
Содержание 3307
Страница 16: ...Steri Cult Incubator ________________________________________________________________________Calibration 3 2 ...
Страница 20: ...Steri Cult Incubator _______________________________________________________________________Configuration 4 4 ...
Страница 21: ...Steri Cult Incubator _______________________________________________________________________Configuration 4 5 ...
Страница 22: ...Steri Cult Incubator _______________________________________________________________________Configuration 4 6 ...
Страница 42: ......
Страница 43: ......
Страница 44: ......
Страница 45: ......
Страница 46: ......
Страница 47: ......
Страница 48: ......
Страница 49: ......
Страница 50: ......
Страница 51: ......
Страница 52: ......