7
•
Specifications
Solvent Monitor
Installation Guide
Page 43
7.2
Physical Specifications
The physical conditions of the device are specified as follows:
Type
Specification
Range of use
Indoor use only
Ambient operating
temperature
5 °C - 35 °C
Ambient storage temperature
-20 °C - 45 °C
Ambient operating humidity
20% - 80% relative humidity (non-condensing)
Ambient storage humidity
Maximum 60% relative humidity (non-condensing)
Operating altitude
Maximum 2000 m above sea level
Pollution degree
2
Power requirements
Via USB connection (max. 500 mA)
Overvoltage category
II
Emission sound pressure level
< 40 dB(A)
Dimensions
(height x width x depth)
13 x 17 x 6 cm
Weight
800 g