Storage Reservoirs
Caution
User-supplied reservoirs must have
an overflow connection to avoid
flooding during the ULTROpure flush
cycle. The overflow must accommo-
date the maximum membrane prod-
uct flow. Failure to provide an ade-
quate overflow will cause flooding
and subsequent equipment and facil-
ities damage.
User-supplied reservoirs are compatible with the ULTROpure.
However, certain precautions must be taken to assure proper
operation of the combination ULTROpure/storage systems.
Care must be used in installing the fill line in the reservoir.
Level sensing switches may be used to control the
ULTROpure. The ULTROpure will accommodate independent
switches for the high and low water levels, or a single high
level switch may be used. In the latter case, the change in
water level between turn on and turn off of the ULTROpure
will be entirely dependent on the mechanical design of the
switch. The Wiring Schematic on page 38 shows the wiring of
the switches.