Utility Connections
A. 6 feet of 1/2" OD tubing is supplied along with
a 1/2" OD x 3/8" NPTM adapter. Wrap Teflon®
tape 1 1/2 to 2 times around this fitting and
install in your incoming feedwater line. Con-
nect the 1/2 OD tubing (refer to Figure C:
Typical Polypropylene Tubing Adapter Installa-
tion) to this fitting and the fitting included on
the ULTROpure. It is recommended that a
customer supplied shutoff valve be installed
on the incoming water line.
B. Connect a sufficient length of 3/8" OD flexible
tubing to the fitting labeled “Reject” on the
right side of the ULTROpure. The unconnect-
ed end should be firmly fastened to an
atmospheric drain. DO NOT RESTRICT THIS
LINE IN ANY WAY.
C. Connect Product line to drain until you finish
flushing the membrane.
Membrane
Installation/Replacement
Installation of the Barnstead reverse osmosis membrane
in the ULTROpure requires the following tools: large slot
blade screwdriver and an open end 5/8" wrench. Install
the membrane by the following steps:
A. If not previously removed, remove top cover
by removing the two screws on the right and
left bottom back corner of the ULTROpure and
pull straight out.
B. Remove the lower cabinet by releasing the
two clamps on both sides of the ULTROpure
and pulling straight out.
Note
If you are replacing an installed mem-
brane, first remove the old membrane
and housing by reversing the following
installation procedure. (i.e. reversing
step L through step A).
I
NSTALLATION