1
Rotor Specifications
Sorvall
Thermo Scientific
75003607 TX-750
1-11
Table 1-10.
Sorvall 100V 50/60Hz rectangular bucket 75003614
Centrifuge
Sorvall Legend XT
Sorvall Legend XF
Sorvall ST 40
Catalog #
75004507
75004534
75004511
Weight empty [kg]
6,8
6,8
6,8
max. cycle number
50,000
50,000
50,000
Maximum permissible load [g]
4x750
4x750
4x750
maximum speed n
max
[rpm]
4,100
4,100
4,100
maximum RCF value at n
max
3,665
3,665
3,665
max. / min. radius [cm]
19.5 / 8.3
19.5 / 8.3
19.5 / 8.9
Pitch angle [°]
90
90
90
Accel. / braking time [s]
65 / 50
65 / 50
65 / 50
sample heating at n
max
[° C] referred to ambient
temperature of 23 °C, running time 120
minutes
8
8
8
aerosol-tight
*
*
Tested by HPA, Porton-down, UK
yes
yes
yes
permissible temperature range autoclavable
°C
121
121
121
Centrifuge
Sorvall Legend XTR
Sorvall Legend XFR
Sorvall ST 40R
Catalog #
75004522
75004540
75004526
Weight empty [kg]
6,8
6,8
6,8
max. cycle number
50,000
50,000
50,000
Maximum permissible load [g]
4x750
4x750
4x750
maximum speed n
max
[rpm]
4,100
4,100
4,100
maximum RCF value at n
max
3,665
3,665
3,665
max. / min. radius [cm]
19.5 / 8.9
19.5 / 8.9
19.5 / 8.9
Pitch angle [°]
90
90
90
Accel. / braking time [s]
65 / 50
65 / 50
65 / 50
aerosol-tight
*
yes
yes
yes
permissible temperature range autoclavable
°C
121
121
121
*
Tested by HPA, Porton-down, UK