4
Rotor Loading
Before a Run
4-2
75003607 TX-750
Thermo Scientific
Before a Run
1. Please read and observe the safety instructions contained in these operating instructions and in
the instructions for use.
2. Check the rotor and all accessory parts for damages such as cracks, scratches or traces of
corrosion.
3. Check the rotor chamber, the motor shaft and the AutoLock
TM
.
4. Check the rotor suitability using the chemical compatibility chart on
page B-1
.
5. Make sure the tubes or bottles do not touch the bucket caps.
6. Check the rotor bolts and apply grease 7500 3786 before using for the first time.
7. Check that each bucket or microplate carrier can swing freely by moving it carefully with your
hand. Weigh the bucket content (adaptor and tube). Make sure you do not exceed the maximum
compartment load.
Proper Loading
To ensure safe operation of the centrifuge, the rotor must be evenly loaded at all times.
You can use two rectangular buckets or two round buckets or two microplate carriers. Make sure that
the opposing receptacles are holding tubes of the same weight.
CAUTION
Always use identical bucket types in a rotor.
Always be sure to use buckets of the same weight class, which is marked on the buckets
themselves.