1
Site Preparation
Power Requirements
Thermo Scientific
TSQ 8000 Preinstallation Requirements Guide
7
• Transient overvoltages must not exceed those specified in category II of IEC
60364-4-443.
2. Use
Table 3
or to determine how many circuits and wall outlets you need. Keep in mind:
• Power must be single-phase.
• Wall outlets must have earth-ground hard-wired to the main panel.
• Included power cords are 2 m (6 ft) long .
Contact your local Customer Service office to discuss power cordset concerns.
3. The TSQ 8000 comes with the required number and type of power cords for your region.
Table 4
will help you identify the correct power cord for your region and instrument. If
you need to replace your power cord or buy an additional cord, you may use any brand of
power cord, as long as it is appropriate for your region.
Table 4.
Power cord identification information by region. (Sheet 1 of 2)
Region
C13 Power Cords (For PC, Monitors,
and Autosamplers)
C19 Power Cords (For GC
and MS)
North American 120V
North American 250V
Japanese 125V
Switzerland 250V
Australian 250V