1
Site Preparation
Entrance Requirements
2
TSQ 8000 Preinstallation Requirements Guide
Thermo Scientific
Entrance Requirements
Use the following guidelines to ensure the entrance to your site will allow for the delivery of
your TSQ 8000 system:
1. Ensure the width of your delivery door opening is at least 81 cm (32 in.).
2. Make sure you have enough room to move boxes around corners, into elevators, or
through doorways. The table below contains the dimensions and weight of shipping
boxes, so that you can make accommodations.
Table 1.
Shipping Box Dimensions and Weight
Box Contents
Length
Width
Height
Weight
cm
in.
cm
in.
cm
in.
kg
lbs.
TSQ 8000 MS
*
107
43
71
28
112
45
138
304
TRACE 1300 or TRACE 1310 GC
60
24
80
31
80
31
64
140
AI/AS 1310 autosampler
36
14
51
20
36
14
10
22
TriPlus RSH (with standard X-axis)
65
26
100
39
32
13
45-55
99-121
TriPlus RSH (with extended X-axis)
65
26
136
54
32
13
60
132
TriPlus 300 Headspace Autosampler
(Box 1 of 2)
77
30
72
28
84
33
70
154
TriPlus 300 Headspace Autosampler
(Box 2 of 2)
79
31
64
25
100
39
25
55
*
The computer, keyboard, monitor, foreline pump, and TSQ 8000 Installation Kit are included in the box with the TSQ 8000.