Thermo Scientific Thermo Fisher 3900 Series Скачать руководство пользователя страница 21

Section 7 Troubleshooting (continued)

 

 
 

 
 

 

Page 7-2 

 

 

TEMPERATURE

 

Temperature too high 

1/ controller  set too high-see Operation section 

 

2/ controller failed on 

– call Technical Services 

Temperature too low 

1/ Over Temperature Limit  Thermostat  set too low 

– see Operation Section 

 

2/ controller set too low 

– see Operation section 

 

3/ unit  not recovered from door opening 

 

4/ unit  not recovered from power  failure  or being turned 

 

5/ controller failure 

– confirm with  front panel lights that controller is operational. 

 

6/ Over Temperature Limit  Thermostat  failure 

– confirm with  front panel lights that thermostat is 

operating correctly. 

 

 

Unit will not heat over a temperature that 

is below set point 

1/ confirm that fan is moving 

– check fan motor motion in shadow box, feel  for air movement in 

chamber 

2/ confirm  that  set point  is set high enough 

–turn  Over Temperature Limit  Thermostat  all  the way 

clockwise  and see if TEC Activated light or OVER TEMP  light  comes on 

 

3/ check calibration 

– using independent certified  thermometer, see Operation section 

 

Unit will not heat/cool at all 

1/ verify that controller is asking for heat/cool  by looking for controller light.  Output 1 should be on 

for heat; output 2 should be on for cooling. 

 

2/ do all controller functions work? 

Indicated  chamber temperature unstable 

1/ ±0.1 may be normal 

 

2/ Are the internal  and external fans working? 

 

3/ Is ambient  room temperature  rapidly changing 

– either door opening or room airflow  from 

heaters or air conditioning? 

– stabilize ambient conditions. 

 

4/ Over Temperature Limit.  Thermostat  set too low 

– thermostat must be set at least one degree 

over the desired main set point; see if OVER TEMPERATURE  pilot  is on continuously; turn controller 

knob completely clockwise  to see if problem solved, then see Operation section for correct setting 

 

5/ electrical  noise-remove nearby sources of RFI including motors, arcing relays, radio transmitters. 

 

Display and reference thermometer  don

’t 

match 

1/ calibration  error 

– see Operation Section 

2/ allow  at least two  hours to stabilize 

Calibrated at one temperature, but not at 

another 

This can be a normal condition  when  operating  temperature  varies widely.  For maximum accuracy, 

calibration  should be done as close to the set point temperature as possible. 

MECHANICAL

 

Motor  makes noise 

Ensure fan is not making contact  with  foreign  material 

Door not sealing 

Confirm  unit  has not  been damaged and unit  is square. Check gasket condition. 

If none of the suggestions listed  above  in the Troubleshooting guide have  solved the problem Technical Services  should 

be contacted  for assistance. Have  the model number, serial number  and voltage (listed  on the date plate on the side of 

the incubator)   as your  service representative  will require it.

 

Содержание Thermo Fisher 3900 Series

Страница 1: ...User Manual Thermo Fisher Model 3900 Series Bod Fly Incubators Operating and Maintenance Manual 7003900 January 2013...

Страница 2: ...volatile or combustible materials are being heated These units have been tested to the requirements of CAN CSA C22 2 No 61010 1 second edition including Amendment 1 or a later version of the same stan...

Страница 3: ...nce must be performed by qualified service personnel Material in this manual is for information purposes only The contents and the product it describes are subject to change without notice Thermo Fish...

Страница 4: ...Temperature Protection Signale un d passement de temp rature Indicates AC Power Rep re secteur AC I Indicates the power is ON I Rep re de la position MARCHE de l interrupteur d alimentation O O Indica...

Страница 5: ...Preface iv...

Страница 6: ...Storing Offset Value for Day Cycle in Memory 4 6 Calibration Offsets for Day Cycle 4 6 Storing Offset Value for Night Cycle in Memory 4 7 Calibration Offsets for Night Cycle 4 7 Settingthe Real Time...

Страница 7: ...es and instructions in this manual may be dangerous and illegal General Safety Considerations Your incubator and its recommended accessories have been designed and tested to meet strict safety require...

Страница 8: ...e data plate at the rear of the unit Do not modify the power cord provided with the unit If the plug does not fit an outlet have a proper outlet installed by a qualified electrician Avoid damaging the...

Страница 9: ...er 2 2 Inspect for concealed loss or damage on the unit itself both interior and exterior If necessary the carrier will arrange for official inspection to substantiate your claim 2 3 Return Shipment S...

Страница 10: ...prevent loss of product due to overloading or circuit failure Note Electrical supply to the unit must conform to all national and local electrical codes 3 2 Location When selecting a site for the inc...

Страница 11: ...d bottom hooks into the provided slots on the sides of the chamber 3 6 Install Shelf Mounts and Shelves Install shelf mounts from the corner top and bottom hooks into the shelf standard in the chamber...

Страница 12: ...e material contained within If temperature rises to the OTL set point the OTL takes control of the peltier unit and prevents the temperature within the incubator from rising further until the problem...

Страница 13: ...ual Process Temperature Display Process Setpoint Display Output 1 Active Heating Element On 20 0C 20 0 1 2 3 EZ1 EZ2 Up Arrow Key Down Arrow Key Output 2 Active Output 3 Active Easy Zone Advance Key L...

Страница 14: ...Event output 2 Event Timer Factory installed timer internal gLbL Global menu HoUr Hour of day night i CA Calibration offset JC Jump Count JL Jump Loop JS Jump Step Min Minutes SoAH Soak oPEr Operatio...

Страница 15: ...LE WILL RUN THE LAST THREE STEPS ARE ENTERING THE NIGHT CYCLE START TIME IN HOURS AND MINUTES AND THE SET POINT WHICH THE NIGHT CYCLE WILL RUN BELOW ARE INSTRUCTIONS ON HOW TO SCROLL THROUGH THE 8 STE...

Страница 16: ...ic thermometer neither are supplied with your incubator in the center of the chamber Placing the reference thermometer in the chamber at this stage of operation will allow for calibrating the control...

Страница 17: ...s been made allow the unit temperature to stabilize before verifying the reference thermometer matches the setpoint Adjust again if necessary 5 5 Setting the Set Over Limit Temperature Control To set...

Страница 18: ...r should be cleaned and disinfected prior to use s 6 1 Cleaning Periodic cleaning is required To clean the incubator perform the following steps 6 1 1 Remove all of the interior parts if assembled 6 1...

Страница 19: ...infectants available that are non corrosive and non abrasive and suitable for use on painted surfaces Contact your local Site Safety Officer for detailed information for the proper disinfectants suita...

Страница 20: ...ify power cord is properly connected there is no loose or disconnected wiring and all control console devices are firmly seated Do not attempt to access the electrical compartment area unless you are...

Страница 21: ...er temperature unstable 1 0 1 may be normal 2 Are the internal and external fans working 3 Is ambient room temperature rapidly changing either door opening or room airflow from heaters or air conditio...

Страница 22: ...14 9660510 9660514 9660510 9660514 9660510 Over Temperature Thermostat OTL 1750813 1750813 1750813 1750813 Pilot Light Green 4650554 4650554 4650554 4650554 Pilot Light Red 4650553 4650553 4650553 465...

Страница 23: ...its 25 5 x 24 0 x 18 5 in 64 77 x 60 96 x 46 99 cm both units 25 5 x 24 0 x 54 5 in 64 77 x 60 96 x 138 43cm both units 25 5 x 24 0 x 54 5 in 64 77 x 60 96 x 138 43cm both units Capacity 6 55 cubic fe...

Страница 24: ...3 4 1 T1 S1 R1 T1 S1 R1 X1 W1 Y1 W Y L3 K3 J3 L4 K4 98 99 CF CD CE B5 D6 D5 WATLOW PM 9660514 RTD SENSOR 6600520 2 3 4 1 2 3 4 1 2 3 4 1 A B C D BROWN BROWN BLUE RED 14G RED 18G BLACK 18G BLACK 14G R...

Страница 25: ...3 4 1 T1 S1 R1 T1 S1 R1 X1 W1 Y1 W Y L3 K3 J3 L4 K4 98 99 CF CD CE B5 D6 D5 WATLOW PM 9660514 RTD SENSOR 6600520 2 3 4 1 2 3 4 1 2 3 4 1 A B C D BROWN BROWN BLUE RED 14G RED 18G BLACK 18G BLACK 14G R...

Страница 26: ...2 L3 K3 J3 L4 K4 98 99 CF CD CE B5 D6 D5 WATLOW PM 9660514 RTD SENSOR 6600520 2 3 4 1 2 3 4 1 2 3 4 1 A B C D BROWN BROWN BLUE RED 14G RED 18G BLACK 18G BLACK 14G RED 22G RED 22G BLACK22G BLACK22G YEL...

Страница 27: ...2 L3 K3 J3 L4 K4 98 99 CF CD CE B5 D6 D5 WATLOW PM 9660514 RTD SENSOR 6600520 2 3 4 1 2 3 4 1 2 3 4 1 A B C D BROWN BROWN BLUE RED 14G RED 18G BLACK 18G BLACK 14G RED 22G RED 22G BLACK22G BLACK22G YEL...

Страница 28: ...R SWITCH 7850571 LED LIGHT BAR 2 3 4 1 T1 S1 R1 T1 S1 R1 X1 W1 Y1 W Y L3 K3 J3 L4 K4 98 99 CF CD CE B5 D6 D5 WATLOW PM 9660510 RTD SENSOR 6600520 2 3 4 1 2 3 4 1 2 3 4 1 A B C D BROWN BROWN BLUE RED 1...

Страница 29: ...3 4 1 T1 S1 R1 T1 S1 R1 X1 W1 Y1 W Y L3 K3 J3 L4 K4 98 99 CF CD CE B5 D6 D5 WATLOW PM 9660510 RTD SENSOR 6600520 2 3 4 1 2 3 4 1 2 3 4 1 A B C D BROWN BROWN BLUE RED 14G RED 18G BLACK 18G BLACK 14G RE...

Страница 30: ...2 L3 K3 J3 L4 K4 98 99 CF CD CE B5 D6 D5 WATLOW PM 9660510 RTD SENSOR 6600520 2 3 4 1 2 3 4 1 2 3 4 1 A B C D BROWN BROWN BLUE RED 14G RED 18G BLACK 18G BLACK 14G RED 22G RED 22G BLACK22G BLACK22G YEL...

Страница 31: ...99 CF CD CE B5 D6 D5 WATLOW PM 9660510 RTD SENSOR 6600520 2 3 4 1 2 3 4 1 2 3 4 1 A B C D BROWN BROWN BLUE RED 14G RED 18G BLACK 18G BLACK 14G RED 22G RED 22G BLACK22G BLACK22G YELLOW ORANGE SSR RELA...

Страница 32: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com...

Отзывы: