Section 5
Operation
Page 5-1
The Peltier module and air circulating fan are used in conjunction with the
temperature control circuit to achieve accurate temperature control. The
thermostat sensor located in the air stream senses any temperature
deviation from the control point. Heating or cooling is provided to
maintain the desired temperature. The circulating fan provides even air
distribution throughout the chamber and assures temperature uniformity.
The power source must match the unit’s requirements listed on the data
plate located on the side of the incubator.
5.1 Plug the service cord into the power source and turn the Power Switch
to the ON position. Turn the Over Temperature Thermostat to its
maximum position, fully clockwise using a coin or flat edged tool.
5.2 Place a liquid filled bottle reference thermometer or the probe of a
remote reading electronic thermometer (neither are supplied with your
incubator) in the center of the chamber. Placing the reference
thermometer in the chamber at this stage of operation will allow for
calibrating the control without the loss of processing time.
5.3 Set Main Temperature: To enter desired setpoint temperature, press
the up/down arrow to raise/lower setpoint. Setpoint is displayed in the
bottom display of the Watlow EZone Controller..
5.4 Calibrating the Main Temperature: It is recommended that calibration
is done once the unit is installed in its working environment. The unit
should be stable at set point for several hours. Once unit has been
stable for several hours, compare process display with reference
thermometer. After comparing the two figures out the calibration offset
by subtracting the reference thermometer reading to the display
reading. If the reference is greater than the display, it will be a positive
number. If the reference is smaller than the display, it will be a negative
number.
EXAMPLE:
Reference thermometer
Process display
Offset cal
27°C
25°C
2
23°C
25°C
-2