Section 6 Maintenance (continued)
Page 6-2
6.2.2. Disinfect the incubator, including all corners, using a suitable
disinfectant. Shelves and shelf clips are autoclaveable. DO NOT USE
spray disinfectants that might leak through openings and cracks and get
on electrical components, or that may contain solvents, corrosives, or
abrasives that will harm the painted coatings. Caution should be used
when cleaning around temperature sensors, and around the door gasket
so as not to impair the positive seal.
6.2.3. If a hazardous material/substance has been damaged in the unit,
immediately initiate your site’s Hazardous Material Spill Containment
protocol. Contact your local Site Safety Officer and follow instructions per
the policy and procedures established for your site.
6.2.4. There are many commercially available disinfectants available that are
non-corrosive and non-abrasive and suitable for use on painted surfaces.
Contact your local Site Safety Officer for detailed information for the
proper disinfectants suitable for your operation.
Warning
Never clean the unit with alcohol or flammable cleaners and
assure all volatile or flammable cleaners are evaporated and dry before
reattaching the unit to the power supply.
Avertissement
Ne jamais nettoyer l'appareil à l'alcool ou avec des
nettoyants inflammables et veiller à ce que les produits volatils ou
inflammables soient entièrement évaporés avant de rebrancher le content
d'alimentation de l'appareil.
No maintenance is required on electrical components. If the incubator fails
to operate as specified, review the TROUBLESHOOTING section prior
to calling for service.
Preventative
Maintenance
Periodically inspect the door latch, trim, catch and gasket for signs of
deterioration. Failure to maintain the integrity of the door system will
shorten the life span of the incubator.