background image

STD5000 & STD6000

Current to Pressure Transducers

User Guide

P/N MA45-5600-00

Revision &

Содержание STD5000

Страница 1: ...STD5000 STD6000 Current to Pressure Transducers User Guide P N MA45 5600 00 Revision...

Страница 2: ......

Страница 3: ...o Fisher makes every effort to ensure the accuracy and completeness of this manual However we cannot be responsible for errors omissions or any loss of data as the result of errors or omissions Thermo...

Страница 4: ...This page intentionally left blank...

Страница 5: ...STD5000 STD6000 User Guide v Revision History Revision Level Date Comments 5F A 06 05 Revised per ECO 4882 B 06 06 Revised per ECO 5219 C 10 07 Revised per ECO 5720 D 08 08 Revised per ECO 6521 E 04...

Страница 6: ...This page intentionally left blank...

Страница 7: ...4 1 General 4 1 Replacing the E Pi 4 2 Fixed Orifice Replacement 4 3 Cleaning Repair 4 4 Calibration the Test Jack 5 1 Chapter 5 Calibration 5 1 The Test Jack 5 1 Calibration Operation 5 1 Monitoring...

Страница 8: ...This page intentionally left blank...

Страница 9: ...nsducer utilizes a virtually weightless membrane that is electromagnetically positioned over a nozzle to precisely modulate the pneumatic backpressure output The low mass membrane is held in a continu...

Страница 10: ...Product Overview Requirements 1 2 STD5000 STD6000 User Guide Thermo Fisher Scientific Figure 1 1 Operation flow chart...

Страница 11: ...le pressure range for calibrated output a Standard supply pressure should be 20 2 psi for 3 15 psig output and 35 2 psi for 3 27 6 30 or 1 17 psig outputs b Supply pressure for standard high pressure...

Страница 12: ...trument housing is designed for mounting to a standard valve yoke 2 inch bolt spacing or a 2 inch pipe A valve mount kit consists of bolts and washers for mounting to the valve yoke A pipe mount adapt...

Страница 13: ...following 1 One inch FNPT port is provided for supply air connections Each unit has a filter screen installed in this port 2 Two inch FNPT ports are provided for pneumatic output connections Either po...

Страница 14: ...The instrument is supplied with two inch FNPT electrical conduit connections Insert a inch FNPT pipe plug into the unused connection Note that the pipe plugs are certified and suitable for the mainten...

Страница 15: ...ignal occurs by measuring the output with a pressure sensor This signal is then compared with the current to the E Pi resulting in exact pneumatic output relative to the input signal Programming Progr...

Страница 16: ...psig SA45 1502 17 SA45 1502 17 SA45 1502 17 4 20 or 10 50 4 12 or 10 30 12 20 or 30 50 35 psig 35 psig 35 psig N A N A N A x x x x x x Zero Zero Zero 17 1 psig 17 1 psig 17 1 psig SA45 1502 17 SA45 1...

Страница 17: ...thus prevent valve operation overlap in split range Table 3 2 shows these values Also note that there is adequate adjustment of span and zero to provide a margin of overlap if desired Table 3 2 Mode...

Страница 18: ...This page intentionally left blank...

Страница 19: ...unit If the replacement E Pi has been in storage for a long period of time it may be necessary to apply a lubricant Dow Corning or equivalent to the O rings Caution When removing or installing the E P...

Страница 20: ...cal connections Press straight down into the manifold and tighten the retaining screw 5 Electrical connections are a two point contact system The first contact occurs when the connector is pushed onto...

Страница 21: ...e fixed orifice with needle nose pliers turn the module over Remove the four screws holding the E Pi to the plastic module It is not necessary to disconnect the wires from the module Do not discard th...

Страница 22: ...t is not necessary to remove them 3 Carefully remove the E Pi top from the base 4 Carefully remove the membrane 5 Insert the straw of an alcohol based contact cleaner or equivalent into the small hole...

Страница 23: ...r to spray off both sides of the membrane Allow the cleaner to air dry Place the membrane on the E Pi base membrane top T side up 7 Use the contact cleaner to spray off the top and allow to air dry 8...

Страница 24: ...This page intentionally left blank...

Страница 25: ...librated range and adjust the zero for 0 of desired output 6 Input 100 of the calibrated range and adjust the span for 100 of desired output 7 Repeat steps 5 and 6 until no further adjustments are nec...

Страница 26: ...turn operation of the unit to the original source by remove the jack plug Figure 5 1 Test jack calibration connections Monitoring 1 Refer to Figure 5 2 and connect an ammeter to the jack plug cable as...

Страница 27: ...nsure reliable long term operation with minimal maintenance An external filter system is available from Thermo Fisher Supply Pressure Regulation 1 Maintain supply air at pressures required by the outp...

Страница 28: ...Pi according to Replacing the E Pi 3 Disconnect the input leads 4 Remove the four screws holding the CCB in the manifold Note The replacement CCB should have an O ring around the pressure sensor mount...

Страница 29: ...ck supply air Unit will not go to full scale with full scale input 1 E Pi is contaminated 2 Circuit board failure 3 3 Leak in tubing 1 Clean or replace Chapter 4 2 Replace circuit board Chapter 6 3 Ch...

Страница 30: ...thin the warranty period Thermo Fisher shall have the right to inspect such products at Buyer s plant or to require Buyer to return such products to Thermo Fisher s plant In the event Thermo Fisher re...

Страница 31: ...rmo Fisher by other suppliers shall carry no warranty except such suppliers warranties as to materials and workmanship Thermo Fisher disclaims all warranties expressed or implied with respect to such...

Страница 32: ...This page intentionally left blank...

Страница 33: ...Thermo Fisher Scientific STD5000 STD6000 User Guide A 1 Appendix A Toxic Hazardous Substances Tables The English and Chinese versions of the Toxic and Hazardous Substances tables are shown below...

Страница 34: ...This page intentionally left blank...

Страница 35: ...5 Other contact Thermo Fisher 6 0 2 1 0 bar standard or high pressure supply Code Case Style 1 Standard H High pressure supply 3 15 psig and 0 2 1 0 bar outputs Code Options 1 Pipe mount kit 2 Test ja...

Страница 36: ...h drive screws Engraved FP82 2022 00 FAS2022 filter regulator with supply gauge GA21 1200 30 Output gauge 0 30 psi CBM 2 inch GA21 1200 60 Output gauge 0 60 psi CBM 2 inch MA45 5600 00 STD5000 STD6000...

Страница 37: ...to CE Conformity for test standards Table C 2 Functional specifications Input 4 20 or 10 50 mA Output 6 30 3 27 3 15 or 1 17 psig and 0 2 1 0 bar Supply Pressure Standard Configuration Minimum of 3 p...

Страница 38: ...ATEX STD5000 Intrinsic Safety 20 C to 60 C STD6000 Intrinsic Safety 20 C to 60 C Flame proof 20 C to 40 C Operational Modes Direct reverse and or split range Failure Mode Transducers always fail in th...

Страница 39: ...and outputs of 3 15 3 27 6 30 and 1 17 psig without Safety Barriers CSA ENC 4 Outdoors Caution Substitution of components may impair Intrinsic Safety Caution Open circuit before removing cover STD600...

Страница 40: ...Specifications Figure C 1 I S installation drawing C 4 STD5000 STD6000 User Guide Thermo Fisher Scientific...

Страница 41: ...emperature The equipment must be connected to a suitably certified ATEX barrier whose parameters must be compatible with the following electrical parameters Ui 30Vdc Ii 100mA Pi 0 75W Ci 5 3nF Li 7 H...

Страница 42: ...with the applicable code of practice e g EN 60079 14 Equipment may also be used in the presence of explosive dust atmospheres in Zones 21 and 22 Installation must be in accordance with the applicable...

Страница 43: ...Specifications Figure C 2 Thermo Fisher Scientific STD5000 STD6000 User Guide C 7...

Страница 44: ...and should not be used outside this range 20 C to 60 C for the STD5000 20 C to 50 C for the STD6000 4 Installation shall be carried out in accordance with the applicable code of practice by suitably...

Отзывы: