RC-5C PLUS
Installation
2-3
C A U T I O N
The centrifuge can be seri-
ously damaged if it is operated when
it is not level.
!
Installation
Install the RC-5C PLUS centrifuge as follows, observing all electrical,
location, and environment requirements listed on the preceding
pages:
1.
If you are installing a new centrifuge,
remove any packaging.
2. Roll centrifuge into position, then turn the two front feet (one at
each corner) to lower them so that they both touch the floor.
3. If the chamber door is not already open, open it by following the
emergency sample recovery procedure in Chapter 5.
4. Install a rotor on the drive spindle WITHOUT its lid. Place a
level on the center hub of the rotor.
5. Using the 9/16-inch wrench provided, alternately turn the two
front feet to raise or lower the front corners of the centrifuge
until the centrifuge is level and both feet bear weight.
Read the
CAUTION/WARNING.
6. Remove the level and the rotor from the rotor chamber.
7. Plug the centrifuge into the appropriate electrical outlet.
2-3/2-4
W A R N I N G
The RC-5C PLUS Centri-
fuge weighs 362 kg (800 lb). Take
the necessary precautions when in-
stalling and moving the centrifuge to
avoid personal injury and/or damage
to the centrifuge.
!
W A R N I N G
The front feet must be used
as explained in this section to lift the
casters off the floor. This will help
minimize centrifuge movement in the
event of a rotor failure.
!