Prepare samples using MicroAmp

 tube strips with

attached caps

1.

Separate the blue tray and retainer by squeezing the release

catch as indicated in the graphic.

1

2

3

1

1

Release catch

2

MicroAmp

 96-Well Retainer

3

MicroAmp

 96-Well Tray

2.

Place the blue tray on the 96-well base.

3.

Load the tube strips on the tray.

4.

Pipette the reaction mixture into the tubes.

5.

Seal the tube strip using a MicroAmp

 cap strip. See “Seal

tubes strips with cap strips“ on page 5 for instructions.

6.

Remove the blue tray containing the sealed tube strips from

the 96-well base and place the tray and sealed tube strips into

the instrument.

1

MicroAmp

 8-Tube Strip with Attached Caps (0.2-mL)

2

MicroAmp

 96-Well Tray

3

MicroAmp

 Splash Free 96-Well Base

Seal tubes strips with cap strips

IMPORTANT! 

Apply significant downward pressure on the

sealing tool in all steps to form a complete seal on top of the tubes.

1.

Align and place the cap strips on the tubes.

2.

Seal the cap strips using the rocking capping tool:

a.

Slip your fingers through the handle with the holes in

the tool facing down.

b.

Place the holes in the tool over the first eight caps in a

row.

c.

Rock the tool back and forth a few times to seal the

caps.

d.

Repeat for remaining caps in the row, then for all

remaining rows.

Prepare samples using MicroAmp

 Reaction Tubes

1.

Set the black tray on a 96-well base.

2.

Place the reaction tubes in the black tray.

1

2

3

1

MicroAmp

 Reaction Tube with Cap (0.2-mL)

2

MicroAmp

 96-Well Tray for VeriFlex

 Blocks

3

MicroAmp

 Splash Free 96-Well Base

3.

Pipette the reaction mixture into the reaction tubes.

4.

Cap the tubes.

5.

Remove the black tray with sealed reaction tubes from the

96-well base and place the tray and sealed tubes into the

instrument.

SimpliAmp

 Thermal Cycler Installation and Operation Quick Reference 

5

Содержание SimpliAmp

Страница 1: ...Set up the SimpliAmp Thermal Cycler IMPORTANT Save the packing materials and box in case you need to ship the instrument in for service 1 Open the shipping crate to unpack the instrument You should re...

Страница 2: ...create an account Link the PCR instrument to Thermo Fisher Connect administrator only The procedure to link the Administrator profile to Thermo Fisher Connect only needs to be performed the first tim...

Страница 3: ...Connect by Thermo Fisher Scientific 2 For Android devices download the application from Google Play by searching for Instrument Connect by Thermo Fisher Scientific 3 Log in to the instrument connect...

Страница 4: ...bes are placed in the sample block without support they can be crushed or deformed when the lid of the thermal cycler is closed over the block Using the tray and retainer set allows the pressure to be...

Страница 5: ...nward pressure on the sealing tool in all steps to form a complete seal on top of the tubes 1 Align and place the cap strips on the tubes 2 Seal the cap strips using the rocking capping tool a Slip yo...

Страница 6: ...V 50 60 Hz Max 600 W VeriFlex blocks 3 VeriFlex zones Supported 10 C 5 C zone to zone Settable 3 20 C 10 C zone to zone Ambient humidity 15 80 Relative Humidity non condensing acceptable range Ambient...

Страница 7: ...Optical 384 Well Clear Reaction Plates with Barcode 4483285 4483273 Strip tubes and caps MicroAmp Fast 8 Tube Strip 0 1 mL 4358293 MicroAmp Optical 8 Tube Strip with Attached Optical Caps 0 2 mL A305...

Страница 8: ...without notice DISCLAIMER TO THE EXTENT ALLOWED BY LAW THERMO FISHER SCIENTIFIC INC AND OR ITS AFFILIATE S WILL NOT BE LIABLE FOR SPECIAL INCIDENTAL INDIRECT PUNITIVE MULTIPLE OR CONSEQUENTIAL DAMAGES...

Отзывы: