Maintenance guidelines

CAUTION! 

During instrument operation, the temperature

of the heated cover can be as high as 110°C, and the

temperature of the sample block(s) can be as high as 100°C.

Before performing the procedure, keep hands away until

the heated cover and sample block(s) reach room

temperature.

To ensure proper operation:

• Regularly:

– Wipe the instrument surfaces with a lint-free cloth.
– Clean the vents, touchscreen, and sample block of the

instrument.

– Clean the sample wells with 100% isopropanol.

• Use only consumables recommended by Thermo Fisher

Scientific for the instrument.

Use of consumables that are larger or smaller than the

specified volume can damage the instrument, contaminate

the sample block, and/or decrease the PCR yield (due to

inefficient thermal transfer).

• Do not use sharp objects on the touchscreen.

Use only your fingers or blunt objects to enter commands on

the instrument touchscreen. Sharp and/or pointed objects

such as writing utensils can damage the surface of the

touchscreen.

• Back up frequently.

Routinely back up the configurations and files on your

instrument to a USB drive. Regular backups protect against

data loss caused by user error, power failure, or instrument

error. For more information, see the SimpliAmp

 Thermal

Cycler User Guide (Pub. No. MAN0009889).

How to use the MicroAmp

 96-well Tray and

Retainer

IMPORTANT! 

Use the MicroAmp

 96-well Tray and Retainer to

prevent crushing of tubes in the thermal cycler.

When small number of tubes (1–2 tube strips or ≤8–16 tubes are

placed in the sample block without support, they can be crushed

or deformed when the lid of the thermal cycler is closed over the

block. Using the tray and retainer set allows the pressure to be

distributed evenly over the tubes and prevents uneven pressure

across the block.
Using the tray and retainer set is optional when using ≥3 tube

strips or ≥16 tubes that are distributed evenly across the block.

• The blue tray and retainer set is compatible for use with

MicroAmp

 single tubes or tube strips with separate cap

strips.

• The blue tray by itself is compatible for use with MicroAmp

tube strips with attached caps.

• The black retainer is compatible for use with single

MicroAmp

 reaction tubes with attached caps.

Prepare samples using MicroAmp

 tubes/tube strips

with separate cap strips

1.

Separate the blue tray and retainer by squeezing the release

catch as indicated in the graphic.

1

2

3

1

1

Release catch

2

MicroAmp

 96-Well Retainer

3

MicroAmp

 96-Well Tray

2.

Place the blue tray on the 96-well base.

3.

Load the tube strips on the tray.

4.

Pipette the reaction mixture into the tubes.

5.

Place the blue retainer over the tubes and snap the retainer

into the tray.

6.

Seal the tube strip using a MicroAmp

 cap strip. See “Seal

tubes strips with cap strips“ on page 5 for instructions.

7.

Remove the blue tray/retainer assembly containing the

sealed tube strips from the 96-well base and place the

assembly into the instrument.

1

2

4

3

5

1

MicroAmp

 8-Cap strip

2

MicroAmp

 96-Well Retainer

3

MicroAmp

 8-Tube Strip (0.2-mL) or MicroAmp

 Reaction

Tube without Cap (0.2-mL)

4

MicroAmp

 96-Well Tray

5

MicroAmp

 Splash Free 96-Well Base

4

SimpliAmp

 Thermal Cycler Installation and Operation Quick Reference

Содержание SimpliAmp

Страница 1: ...Set up the SimpliAmp Thermal Cycler IMPORTANT Save the packing materials and box in case you need to ship the instrument in for service 1 Open the shipping crate to unpack the instrument You should re...

Страница 2: ...create an account Link the PCR instrument to Thermo Fisher Connect administrator only The procedure to link the Administrator profile to Thermo Fisher Connect only needs to be performed the first tim...

Страница 3: ...Connect by Thermo Fisher Scientific 2 For Android devices download the application from Google Play by searching for Instrument Connect by Thermo Fisher Scientific 3 Log in to the instrument connect...

Страница 4: ...bes are placed in the sample block without support they can be crushed or deformed when the lid of the thermal cycler is closed over the block Using the tray and retainer set allows the pressure to be...

Страница 5: ...nward pressure on the sealing tool in all steps to form a complete seal on top of the tubes 1 Align and place the cap strips on the tubes 2 Seal the cap strips using the rocking capping tool a Slip yo...

Страница 6: ...V 50 60 Hz Max 600 W VeriFlex blocks 3 VeriFlex zones Supported 10 C 5 C zone to zone Settable 3 20 C 10 C zone to zone Ambient humidity 15 80 Relative Humidity non condensing acceptable range Ambient...

Страница 7: ...Optical 384 Well Clear Reaction Plates with Barcode 4483285 4483273 Strip tubes and caps MicroAmp Fast 8 Tube Strip 0 1 mL 4358293 MicroAmp Optical 8 Tube Strip with Attached Optical Caps 0 2 mL A305...

Страница 8: ...without notice DISCLAIMER TO THE EXTENT ALLOWED BY LAW THERMO FISHER SCIENTIFIC INC AND OR ITS AFFILIATE S WILL NOT BE LIABLE FOR SPECIAL INCIDENTAL INDIRECT PUNITIVE MULTIPLE OR CONSEQUENTIAL DAMAGES...

Отзывы: