2. Position the platform mounting plate atop the shaker body and install
all 9 screws by hand until they are finger-tight only. Rotate the speed
control fully counterclockwise (OFF) and plug the unit into an outlet.
Rotate the speed control knob slowly in the clockwise direction to
allow the shaker to orbit at its slowest speed.
Note
While adjusting the platform retaining screws, make certain lid
switch, located at back of chamber, is activated (pressed in).
3. While it is slowly orbiting, slightly tighten each of the A screws in
Figure 1. Repeat this procedure by slightly tightening all the B screws,
then complete procedure by tightening the C screws.
Caution
All screws must first be slightly tightened to seat plate properly.
s
4. Repeat the screw-tightening procedure at 50-100 RPM until all 9
screws are securely tightened.
1. Referring to
refrigerated models SHKE6000-7 and SHKE6000-8CE
only, first disconnect power cord.
2. Remove top 3 screws from control panel nose
3. Remove air filter and clean. Reinstall air filter.
4. Replace control panel nose onto unit. Make sure the studs located on
the bottom flange go through holes located on the bottom of the nose.
5. Replace and tighten the top 3 screws for the control panel nose.
6. If needed remove condenser side cover (left side of unit) and inspect
condenser coil. If needed, clean with mild soap and water, rinse off
with clean water and dry thoroughly. Take special care not to bend
condensor fins.
7. Replace condenser side cover.
Suggested with every 3 months of constant use
Any internal adjustments or repairs must be performed by a qualified
service representative.
Remove the bearing pan by loosening 4 thumbscrews in the platform
center. Remove the sheet metal panel (9 screws) under the platform to
expose the belt and interior parts. Inspect the drive belt for wear. Order a
replacement if necessary.
6-2
MaxQ 6000 Shaker
Thermo Scientific
Section 6
Maintenance
Replacing Platform
Mntg Plate (cont.)
Cleaning Condenser
Filter / Coil
Platform
Maintenance
Содержание SHKE6000 (4359)
Страница 34: ...MaxQ 6000 Shaker 8 1 Thermo Scientific...
Страница 35: ...8 2 MaxQ 6000 Shaker Thermo Scientific...