MaxQ 6000 Shaker 7-3
Thermo Scientific
Section 7
Troubleshooting
Shaking Error Codes
Displayed
Message
Intended to Detect
Cause
Solution
E12
Locked rotor in shaking
control.
Locked rotor condition.
Free locked rotor.
Failure in motor.
Replace motor.
Failure in motor control
circuit.
Replace PC1415X1 control
board.
E13
Shaker RPM too far
from set point.
Shaker pan adjusted
incorrectly.
Readjust the 9 screws
that secure the pan.
Shaking RPM deviates
more than ±10% of set
point.
Shaking platform load
too large.
Obstruction between
shaking platform and
chamber wall.
Remove obstruction.
‘bAL’
Shaker balance error.
The speed display will
show ‘bAL’.
Unit vibrates and pro-
duces excessive noise
and then displays ‘bAL’.
Unit not leveled properly.
Shaking platform load
too large
Reduce the load placed
on shaker platform.
General Error Codes
Displayed Message Intended to Detect
Cause
Solution
E21
Corrupt Data Flash
Memory
Invalid Data Flash
Checksum
Call Technical
Services or see Notes
on E21 below.
Continuous Beep, No
Display.
Corrupt Program Flash
Memory
Invalid Program Flash
Checksum
Replace or Reprogram
PC1415X1 control
board. Call Technical
Services.