1
Safety Information
Introduction
Thermo Fisher Scientific instruments are designed for convenient and reliable service; however, improper use or handling by
a user may damage the instrument, or cause a hazard to health. The instrument must not be used in a manner not specified
by Thermo Fisher Scientific. Correct maintenance procedures are essential for consistent performance. It is recommended
that users secure a maintenance contract with our service department.
Any problems and queries should be referred to your Thermo Fisher Scientific service department.
The following sections contain important information for the safe setup and use of the
instrument, and should be read and understood by the user before using the instrument.
General Safety
This instrument, as supplied, conforms to IEC 61010-1; however, the addition of
chemicals introduces potential hazards. Good Laboratory Practice must be
employed and consideration must be given to the potential for hazard when
dealing with these chemicals.
Do not use the instrument in close proximity to strong electromagnetic radiation,
as these may interfere with the proper operation. The electromagnetic
environment should be evaluated prior to operation of the device.
Do not introduce any source of ignition into, or near, the instrument once it has
been loaded with reagents.
Do not remove any panels or access covers, unless specifically instructed to do so.
The instrument does not have any user serviceable parts. Potentially lethal voltages
are present inside the instrument.
The instrument must be properly connected to a good earth (ground) via the
Mains input supply and positioned such that it is possible to interrupt the Mains
supply at the source by removing the plug from the socket.
Use only factory approved accessories or replacement parts within the instrument.
Only use reagents recommended in the operator guide.
Position the instrument such that it is possible to interrupt the Mains supply at
the source by removing the plug from the socket.
If the equipment is used in a manner not specified by Thermo Fisher Scientific,
the protection offered by the equipment may be impaired.
Environment
This instrument is required to comply with the European Union's Waste Electrical and Electronic Equipment (WEEE)
Directive 2002/96/EC. It is marked with the following symbol:
Содержание PrintMate AS
Страница 1: ...Thermo Scientific PrintMate AS Operator Guide A83810100 Issue 2...
Страница 4: ......
Страница 8: ......
Страница 12: ......
Страница 14: ......
Страница 15: ...7 Chapter 1 Introducing PrintMate AS Identification of Parts System Interaction System Specification...
Страница 17: ...Chapter 1 Introducing PrintMate AS 9 PrintMate AS external connections PrintMate AS Print Head...
Страница 24: ...PrintMate AS A83810100 Issue 2 16 Remove the foam tray Remove the foam base mat and the operator pack...
Страница 36: ......
Страница 54: ......
Страница 76: ......
Страница 77: ...69 Chapter 5 Cleaning and Maintenance Cleaning and Maintenance Cleaning Procedure Changing Fuses...
Страница 82: ......