Chapter 4 - Troubleshooting
57
Instrument Not Responding
Error Message / Symptom
Possible Cause
Remedy
Instrument not responding
Software error.
Check connections.
Switch off PrintMate AS, wait 5
seconds then switch back on.
Wait until the Touch Panel
Display says PrintMate AS
Ready before resuming
operation.
Close down the LabWriter
Application and restart it.
Switch off PrintMate AS.
Close down the LabWriter
Application.
Reboot the PC.
Restart the LabWriter
Application and PrintMate AS.
Check that the COM Port
Settings for LabWriter are
correct.
Check that the Baud Rate
settings for LabWriter are
correct:
PrintMate 19200
PrintMate AS 57600