Operator Guide
59
Lower Line of Print is Misaligned
Error Message / Symptom
Possible Cause
Remedy
Print Template set up incorrectly.
Ask Administrator to check
Print Template setup.
Use Print to Screen
function to check print
layout.
Tape jammed.
Open Print Door and Waste
Spool Access.
Use Tape Feed function on
Touch Panel to check that
Tape feeds correctly.
If Tape does not feed
remove and refit Tape (see
Chapter 2 - Changing the
Foil Tape).
Cassette Clamp error.
Open Print Door and print
a Cassette.
Check Cassette is held
securely during printing and
does not move until the
Cassette Clamp releases it.
If movement is seen call the
Thermo Fisher Scientific
Service Department.
Print Head obstructed.
Open Print Door and check
under Cassette Clamp for
debris.
Incompatible Cassette.
Check that Cassette type is
included on Appendix C -
Approved Cassette List.
Lower line of print is misaligned
Instrument Malfunction.
Call the Thermo Fisher
Scientific Service
Department.