Dispense Lever Fails to Reset
Error Message / Symptom
Possible Cause
Remedy
Cassette blockage when exiting
Hopper.
Open the Print Door and
remove any visible Cassette.
Use the Touch Panel or PC
to continue.
Cassette blockage in Hopper.
Remove the Hopper at the
output position and use the
Touch Panel or PC to
perform a Dispense Cycle
and reset the dispense lever.
Remove the Cassette Tube
and / or any loose Cassettes
in the Hopper.
Manually open and close
the Hopper Slider a few
time to ensure it moves
freely.
Carefully reload the
Cassettes and refit the
Hopper to the Carousel.
Use the Touch Panel or PC
to continue.
Cassette not compatible.
If using a non-approved Cassette,
check that they manually dispense
properly.
Other Hopper Malfunction
Fit another Hopper in the
output position and check
that it operates correctly.
If it does, this indicates that
either the Cassettes or the
Hopper is faulty.
Dispense lever fails to reset
Instrument Malfunction
Call the Thermo Fisher Scientific
Service Department.