background image

Semidry Electroblotter

4-1

Thermo Scientific

Section 4  

Troubleshooting

Problem

Cause

Solution

Transfer efficiency is poor

• Current is too low

• Semidry transfer should be performed at constant

current. Current density should be between 0.5 and
3mA/cm2 of stack surface area.

• Power supply is inappropriate for semidry

transfer

• Many power supplies will shut off or blow a fuse

when run at the conditions required for semidry
transfer. Semidry transfer requires low voltage (often
less than 10V) and high current. Check with the man-
ufacturer of the power supply to determine whether
it is appropriate for semidry transfer.

• Transfer performed for too short a time

• Increase the amount of time for transfer.

• Transfer sandwich was assembled in the

wrong order

• The semi-dry electroblotter is configured with the

cathode on the bottom, and anode on top. This
means that an upward transfer is being performed
rather than downward. Follow the instructions care-
fully when assembling the transfer sandwich.

• The pH of the transfer buffer is too close

to the isoelectric point of the protein

• Try a more acidic or basic transfer buffer.

• Too much methanol in the transfer buffer

• Reducing methanol can help elute proteins from

the gel, but can reduce binding to nitrocellulose
membranes.

• High percentage gels restrict transfer

• Higher percentage acrylamide or crosslinker can

restrict elution of proteins. Use the lowest percent-
age acrylamide possible to separate your proteins.

• Puddles of buffer were present on the

cathode allowing the current to bypass the
stack

• Always clean up the lower plate before closing

the lid of the transfer apparatus. Do not squeeze the
stack excessively as this also creates puddles that
the current can pass through.

• The filter paper was too dry

• Filter paper should be saturated with transfer

buffer before adding them to the sandwich.

Содержание OWL HEP-1

Страница 1: ...Semidry Electroblotter Models HEP 1 and HEP 3 Operating and Maintenance Manual 7007332 Rev 0 Visit us online to register your warranty www thermoscientific com warranty...

Страница 2: ...Thermo Scientific MANUAL NUMBER 7007332 0 5 1 12 Transfer to Marietta was The Panther 3 2003 ccs REV ECR ECN DATE DESCRIPTION By Preface Semidry Electroblotter i...

Страница 3: ...ngs found on the lower buffer chamber s This system is designed to meet IEC 1010 1 safety standards IEC 1010 1 is an internationally accepted electrical standard for laboratory instruments Statement o...

Страница 4: ...nt that has been put on the market after 13 August 2005 This product is required to comply with the European Union s Waste Electrical Electronic Equipment WEEE Directive 2002 96 EC It is marked with t...

Страница 5: ...can fill your needs for spare or replacement parts or provide you with on site service We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products Whatever Th...

Страница 6: ...wich Assembly 2 4 Using the System 3 1 Running the Blot 3 1 Transfer settings 3 1 Running Conditions 3 2 Factors That Affect Transfer Efficiency 3 2 Troubleshooting 4 1 Technical Tips 5 1 Semidry Blot...

Страница 7: ......

Страница 8: ...nsfer at low voltages without external cooling systems Plate electrodes also provide a uniform electric field for efficient even transfers Before starting unpack the unit and inventory your order If a...

Страница 9: ...Heavy Duty Knobs HEP 1 3 HEP 3 4 Power Supply Leads Lid Anode Cathode Base Figure 1 1 Exploded View Table 1 1 Parts List 1 base with stainless steel cathode 1 lid with platinum titanium anode 1 attac...

Страница 10: ...modifiers often use in protein transfer buffer These components however are antithetical in their effects both in terms of movement and adsorption Methanol restricts protein movement from the gel but...

Страница 11: ...l transfer protocol following separately DNA RNA If these gels were not run in IXTBE they should be equilibrated for 10 minutes in this buffer Protein Gels After electrophoresis remove the gel assembl...

Страница 12: ...f the bottom plate In this case flip the gel sandwich over and follow the same procedure 4 Once the plates are separated remove the second side spacer along with any extraneous bits of acrylamide arou...

Страница 13: ...clipping a comer of the membrane or using a ball point pen Clip the same comer until you retire 3 Wet the membrane according to its manufactures recommendations followed by a quick equilibration in t...

Страница 14: ...the filter stack 9 Add a few mL of buffer to the gel and gently layer the membrane as you did the gel 10 Repeat with three more pieces of filter paper 11 Holding the stack drain off all excess buffer...

Страница 15: ......

Страница 16: ...transfer down to say coincide with the setting up of a probe simply decrease the current mA to match the added time you require mA hr Std setting mA hr New setting Alternatively the current can be inc...

Страница 17: ...r efficient transfer of a specific molecule Factors That Affect Transfer Efficiency Protein HEP 3 Sequencing gel see HEP 1 set tings for other size gels on this unit DNA RNA Filter Paper FP 1 20 x 20c...

Страница 18: ...m and anode on top This means that an upward transfer is being performed rather than downward Follow the instructions care fully when assembling the transfer sandwich The pH of the transfer buffer is...

Страница 19: ...Alcohol was not used to prewet the membrane PVDF is hydrophobic and requires a short soak in methanol prior to transfer Air spaces are interfering with contact between the gel and the membrane Roll a...

Страница 20: ...delines are HEP l 2mA cm2 of gel for 1 hour These guidelines are just a starting point and exact conditions have to be determined Different kinds of blotting Western Blotting is a blotting method for...

Страница 21: ...Eckerskorn Christoph and Lottspeich Friedrich Structural characterization of blotting membranes and the influence of membrane parameters for electroblotting and subsequent amino acid sequence analysi...

Страница 22: ...with a useful reference list It also includes a very useful troubleshooting guide for nucleic acid and protein blots with pictures of the problems description of symptoms and proposed solutions 1X Tr...

Страница 23: ...Electroblotting This buffer is used with the HEP 1 Semidry Electroblotter Final 1 X composition Anode 1 Buffer 0 3M Tris Base 20 MeOH pH 10 4 Anode 2 Buffer 0 025M Tris Base 20 MeOH pH 10 4 Cathode B...

Страница 24: ...0 8M Tris 1 18M borate 24mM EDTA pH 8 3 CAPS Buffer pH 11 This buffer can be used to improve transfer of some proteins 10mm CAPS 3 cyclohexylamino 1 1 0 methanol propanesulfuric acid adjust to pH 11 w...

Страница 25: ......

Страница 26: ...use ethanol or other organic solvents to clean these products Do not autoclave bake or microwave your unit Temperatures over 50 C can damage acrylic s Note If an RNase free electrophoresis system is d...

Страница 27: ...S Phenol 5 solution U Hydroxide 10 S Glycerine Heptane commercial grade S Soap solution Ivory S Ammonium Hydroxide concentrate S Hexane S Sodium carbonate 2 S Aniline D Hydrochloric acid 10 S Sodium...

Страница 28: ...f 100 FP 3 Blotting Filter Paper 9cm x 9cm pkg of 100 FP 4 Blotting Filter Paper 10cm x 10cm pkg of 100 FP 6 Blotting Filter Paper 12cm x 16cm pkg of 100 FP 7 Power Supply Leads PSL 5 Buffer Kit recom...

Страница 29: ......

Страница 30: ...t beyond the original warranty period The Technical Services Department must give prior approval for return of any component or equipment At Thermo s option all non conforming parts must be returned t...

Страница 31: ...or to the component part beyond the original warranty period The Technical Services Department must give prior approval for return of any component or equipment At Thermo s option all non conforming p...

Страница 32: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com...

Отзывы: