• If the resistance values are not out of expected range, but electrode response continues to be slow or
erratic, these effects may also be seen as a result of reference clogging or contamination. You should
reference the additional electrode care notes below.
• If the resistance value between checks increases substantially higher, the fill solution may be
depleted. Refer to the additional electrode care notes below for the proper fill solution levels of a
refillable electrode.
• If the resistance values show a “super high” resistance, cleaning and care may improve performance,
but will not be able to return the probe to like new condition. Consider purchasing a new electrode.
Electrode Diagnostics Log
Electrode SN
Date
Temperature
Resistance
Delta*
Notes
SN 1234
(example)
1/1/10
350
–
Initial test
SN 1234
(example)
1/8/10
20
320
Significantly low
resistance indicates
electrode failure.
* Change in resistance value from prior reading.
Содержание Orion Versa Star VSTAR-PH
Страница 1: ...Thermo Scientific Orion Versa Star pH with LogR Temperature Module Reference Guide...
Страница 6: ...This page intentionally left blank...
Страница 12: ...8 Thermo Scientific Orion Versa Star pH with LogR Temperature Module This page intentionally left blank...
Страница 18: ...14 Thermo Scientific Orion Versa Star pH with LogR Temperature Module This page intentionally left blank...
Страница 20: ...16 Thermo Scientific Orion Versa Star pH with LogR Temperature Module This page intentionally left blank...
Страница 38: ...34 Thermo Scientific Orion Versa Star pH with LogR Temperature Module Notes...
Страница 39: ...This page intentionally left blank...