17
Thermo Scientific Orion
V
ersa
s
tar
pH with LogR Temperature Module
Chapter 8 Data Storage and Review
Data Storage Settings
Measurement Read Type
The measurement read type determines when the meter sends measurements to outputs as selected in the setup menu
and when properly connected to the meter. These are set individually for each channel. Measurement Read Type is set
per channel in the setup menu. Please refer chapter 3 Setup Menus in this reference guide for more information.
Datalog, Computer and Printer Export Settings
Orion
V
ersa
s
tar
Meters have a 2000-datalog capability. To make your selection:
1. From measurement mode, press
setup/3
.
2. Press the arrow keys to highlight Inst. Settings. Press
setup/3
. Use the
down arrow
key to highlight
your desired communication output port:
a. USB
b. RS232
c. USB Printer – enables communication through the micro-USB port directly to a USB printer.
3. Press
f2 (page)
key, use the
arrow
keys and
f3 (select)
key to:
a. turn on printing
b. select the print format
c. turn on datalogging
For more information on these settings, refer to the Orion
V
ersa
s
tar
Reference Guide, Chapter 3 Instrument Settings to
turn on Data Log to enable data storage.
Datalog Review
This option allows for stored data to be:
• reviewed
• printed individually, in a group or all
• deleted
• analyzed with general statistical information
The datalog will reflect a screen capture of all the information at the time of datalogging.
To review datalog information
• Press
log view (9)
key
• Press
f3 (select)
key
• Follow prompts to review information
Содержание Orion Versa Star VSTAR-PH
Страница 1: ...Thermo Scientific Orion Versa Star pH with LogR Temperature Module Reference Guide...
Страница 6: ...This page intentionally left blank...
Страница 12: ...8 Thermo Scientific Orion Versa Star pH with LogR Temperature Module This page intentionally left blank...
Страница 18: ...14 Thermo Scientific Orion Versa Star pH with LogR Temperature Module This page intentionally left blank...
Страница 20: ...16 Thermo Scientific Orion Versa Star pH with LogR Temperature Module This page intentionally left blank...
Страница 38: ...34 Thermo Scientific Orion Versa Star pH with LogR Temperature Module Notes...
Страница 39: ...This page intentionally left blank...