Chapter 5  Temperature Calibration

For improved temperature accuracy, the Orion 

V

ersa

 s

tar

 meter and pH with LogR Temperature module system offers 

the option for temperature calibration. One point can be calibrated for an Automatic Temperature Compensation (ATC) 
probe. Up to three points can be calibrated when LogR temperature input is being used. Please review the information 
below prior to calibration.

LogR Temperature Calibration Information

This procedure should be performed at least once per week to ensure accurate temperature compensation when 
operating in the LogR mode. If the meter is powered down or the electrode is disconnected, the LogR temperature 
should be recalibrated. When using the meter for the first time or restarting the meter in the LogR mode, a LogR 
temperature calibration must be performed. It is recommended that at least a two-point temperature calibration be 
performed to meet optimum system accuracy. Up to three points can be calibrated. Refer to the 

Appendix B

 for 

additional information. 

Choose a temperature calibration range that is at least 5 ºC apart, depending on your expected operating range.  
Evenly bracket your expected sample temperature range. A maximum 20 ºC range is recommended for maximum 
accuracy. For example, for measurements around 20 ºC, calibrate at 10 ºC and 30 ºC. For a temperature span greater 
than 20 ºC, a three-point temperature calibration is recommended. For many samples, a single-point calibration near the 
expected sample temperature will yield adequate results. A constant temperature bath or a hot plate is recommended 
to hold the temperature of calibrating solution. 

For best results during temperature calibration, it is recommended that a pH buffer or tap water be used as  
the temperature calibrating solution. Do not use distilled water because the conductivity is too low to obtain a  
good calibration.

LogR Calibration Technique

LogR calibration can be done by using an ATC probe or with a NIST reference. An accurate and the fastest method is to 
use an ATC probe. The most accurate method is to use a NIST reference, such as a NIST-traceable thermometer. 

Note:

 The meter defaults to using an ATC probe for LogR calibration. If the temperature reading for the ATC probe 

requires adjustment, follow the instructions for ATC probe calibration technique in the next section.

1.   In the measurement mode, press 

setup/3 

key.

2.   Use the

 arrow

 keys to highlight the pH channel with LogR and press 

f3 (Select)

.

3.   Press the 

right arrow

 key to highlight Temperature, and press 

f3 (Select)

.

4.  

Temperature Input 

should be set to LogR. Use the 

arrow

 keys to highlight the appropriate LogR 

temperature channel if LogR is not already selected. Select the appropriate channel with LogR for 

Temperature Cal

 and press 

f3 (Select)

.

Содержание Orion Versa Star VSTAR-PH

Страница 1: ...Thermo Scientific Orion Versa Star pH with LogR Temperature Module Reference Guide...

Страница 2: ...riode Minimum Stir Requirement MSR NISS Optimum Results Orion Dual Star Orion Star SAOB SMART AVERAGING SMART STABILITY Star LogR Star Navigator 21 Stat Face Triode are trademarks of Thermo Fisher Sci...

Страница 3: ...libration 9 pH Calibration 9 RmV Calibration 9 ORP Calibration 10 Chapter 5 Temperature Calibration 11 LogR Temperature Calibration Information 11 LogR Calibration Technique 11 ATC Probe Calibration T...

Страница 4: ...Appendix B About LogR Technology LogR Temperature Measurements 29 LogR Temperature Accuracy Tables 29 Appendix C Advanced Orion Versa Star Meter and pH with LogR Module s Electrode Diagnostics Electr...

Страница 5: ...01 and 0 1 mV and RmV ranges from 2000 mV Automatic and manual datalogging options Up to 6 point pH calibration with automatic recognition for USA NIST and DIN buffers The ability to fix calibration...

Страница 6: ...This page intentionally left blank...

Страница 7: ...assign the probe to the channel in the setup menu for instrument settings 4 Set up work area with calibration buffers rinse water sample other supplies 5 Connect the meter for data collection Hyperter...

Страница 8: ...dition and slope Calibration information with electrode slope from last calibration and buffer points The information on the display can be customized Enter the setup menu to make changes 1 Press Setu...

Страница 9: ...Stability Smart Fast Medium Slow Averaging Off Automatic Smart Read Type AutoRead Timed Single Shot Continuous Alarm Limit CalDue Set Point Note Different modes mV RmV ORP in the pH module have setup...

Страница 10: ...he meter will apply the same saving and exporting data conditions as with the set time interval to record and export data but lock after one timed reading has expired Press measure esc 1 to reset the...

Страница 11: ...annel with LogR and press f3 Select 3 Press the right arrow key to highlight Temperature and press f3 Select 4 Use the arrow keys to highlight the desired Temperature Input The selection is automatica...

Страница 12: ...8 Thermo Scientific Orion Versa Star pH with LogR Temperature Module This page intentionally left blank...

Страница 13: ...setup menu and the meter allows custom buffers to be used for up to 6 calibration points Custom buffer option prompts for the nominal values and stores data 4 During calibration points can be edited o...

Страница 14: ...MD and Orion 4M KCl filling solution cat no 900011 or 4M KCl gel in electrode with cat no 9179BNMD are used The raw mV reading of an electrode with the same conditions is 220 mV at 25 C One standard c...

Страница 15: ...y For example for measurements around 20 C calibrate at 10 C and 30 C For a temperature span greater than 20 C a three point temperature calibration is recommended For many samples a single point cali...

Страница 16: ...Done key must be pressed Press measure esc 1 key or f1 Back to exit calibration without saving ATC Probe Calibration Technique The ATC temperature display has a relative accuracy of 0 1 C ATC probes...

Страница 17: ...key If the meter is in timed or continuous read modes the meter will immediately start taking readings 4 If in AutoRead mode and datalog is enabled default the reading will automatically be stored whe...

Страница 18: ...14 Thermo Scientific Orion Versa Star pH with LogR Temperature Module This page intentionally left blank...

Страница 19: ...he last calibration data stored for each method Use the channel setup to access and modify methods 1 Press f3 setup key 2 Use the arrow keys to highlight the appropriate channel Press f3 select 3 Use...

Страница 20: ...16 Thermo Scientific Orion Versa Star pH with LogR Temperature Module This page intentionally left blank...

Страница 21: ...ys to highlight Inst Settings Press setup 3 Use the down arrow key to highlight your desired communication output port a USB b RS232 c USB Printer enables communication through the micro USB port dire...

Страница 22: ...of the last 10 calibrations for each parameter of the meter The information will match the information recorded for that calibration To review calibration information 1 Press log view 9 key 2 Press ri...

Страница 23: ...roubleshooting Guide Assistance After troubleshooting all components of your measurement system contact Technical Support Within the United States call 1 800 225 1480 and outside the United States cal...

Страница 24: ...A Equipment Class Measurement control and laboratory Orion Star A series meters are EMC Class A Conforms with the following directives and standards EN61326 1 2006 Electromagnetic Compatibility EMC Di...

Страница 25: ...EH ORP Mode Yes Temperature Range 5 to 105 C 23 to 221 F Resolution 0 1 Relative Accuracy 0 1 Offset Calibration 1 point LogR Temperature Range 0 to 100 C 0 to 212 F Resolution 0 1 Relative Accuracy...

Страница 26: ...meter Orion Versa Star conductivity temperature module Electrode arm with redesigned holder Universal power adapter VSTAR22 Orion Versa Star Conductivity Benchtop Meter Kit Orion Versa Star meter Orio...

Страница 27: ...with Two pH ISE Modules Kit Orion Versa Star meter Two Orion Versa Star pH ISE temperature modules 8102BNUWP Orion ROSS Ultra pH electrode 096019 Orion Star stirrer probe 927007MD ATC probe stainless...

Страница 28: ...trode storage bottle 810007 ROSS fill solution 5x60 mL Electrode arm with redesigned holder Universal power adapter VSTAR90 Orion Versa Star pH ISE Conductivity RDO Dissolved Oxygen Benchtop Multipara...

Страница 29: ...electrode storage bottle 011007 conductivity standard 1413 S 5x60 mL Electrode arm with redesigned holder Universal power adapter VSTAR93 Orion Versa Star pH ISE Conductivity RDO Dissolved Oxygen Ult...

Страница 30: ...body 927005MD Orion ATC probe epoxy body 013005MD Orion DuraProbe conductivity cell 4 cell K 0 475 013016MD Orion conductivity cell flow through K 0 1 083005MD Orion dissolved oxygen probe polarograp...

Страница 31: ...n use is capable of performing automatic buffer recognition perform the following procedure 1 Set the measurement mode to mV Prepare the pH electrode according to the instructions in the pH electrode...

Страница 32: ...int calibrations in the pH mode In this technique straight line segments are drawn between each successive pair of calibration points Separate slope and E0 values are calculated using the adjacent pai...

Страница 33: ...logy makes this possible using most standard glass pH electrodes The system is based on using the electrical resistance of the glass sensing bulb as the temperature probe The logarithm of the resistan...

Страница 34: ...N 0 29 0 06 0 003 0 001 9207BN 0 16 0 07 0 002 0 001 9256BN 0 27 0 07 0 003 0 001 9272BN 0 22 0 06 0 003 0 001 For 20 C temperature compensation spans Table 2 Three Point LogR Calibration Average Temp...

Страница 35: ...maintenance will not be able to rejuvenate your electrode Viewing Electrode Resistance The pH electrode bulb resistance value can be found right on the measurement screen when LogR is selected for th...

Страница 36: ...values in the table below While diagnostic testing at or near 25 C is ideal due to the resistance stability at this temperature this may not be possible Nonetheless it is important to maintain as many...

Страница 37: ...the fill solution may be depleted Refer to the additional electrode care notes below for the proper fill solution levels of a refillable electrode If the resistance values show a super high resistanc...

Страница 38: ...34 Thermo Scientific Orion Versa Star pH with LogR Temperature Module Notes...

Страница 39: ...This page intentionally left blank...

Страница 40: ...thermo com India Tel 91 22 4157 8800 wai asia thermofisher com Japan Tel 81 045 453 9175 wai asia thermofisher com China Tel 86 21 68654588 wai asia thermofisher com Singapore Tel 65 6778 6876 wai asi...

Отзывы: