Error
Recommended Actions
Measurement is
flashing 9999 and over
range or under range.
Measurement is outside the allowable measurement range. Check electrode
connection and settings in the setup menu. Clean the electrode according to the
electrode user guide and recalibrate the electrode with new buffers or standards.
If the error continues, perform the meter self test in the diagnostics menu.
The measurement
freezes and will
not change.
The meter is in AutoRead mode. (AR icon appears solid on the display.)
Press
measure (esc)/1
key to take a new reading or go to setup to change the
read type.
The meter does not
automatically recognize
the pH buffer
during calibration.
Verify the correct buffer set was selected in the setup menu. The meter uses
raw mV readings to recognize the buffer. As the electrode ages or becomes dirty,
its mV readings will drift. Check buffers and clean electrode according to
electrode instructions.
Chapter 9 Customer Services
Troubleshooting Guide
Assistance
After troubleshooting all components of your measurement system, contact Technical Support. Within the United
States call 1.800.225.1480 and outside the United States call 978.232.6000 or fax 978.232.6031. In Europe, the Middle
East and Africa, contact your local authorized dealer. For the most current contact information, or the latest application
and technical resources for Thermo Scientific Orion products, visit www.thermoscientific.com/water.
Warranty and Registration
To register your new meter and for the most current warranty information, visit www.thermoscientific.com/water.
WEEE Compliance
This product is required to comply with the European Union’s Waste Electrical & Electronic Equipment (WEEE) Directive
2002/96/EC. It is marked with the symbol above.
Thermo Fisher Scientific has contracted with one or more recycling/disposal companies in each EU Member State and
this product should be disposed of or recycled through them. Further information on compliance with these directives,
the recyclers in your country, and information on Thermo Scientific Orion products that may assist the detection of
substances subject to the RoHS Directive are available at www.thermoscientific.com.
Содержание Orion Versa Star VSTAR-PH
Страница 1: ...Thermo Scientific Orion Versa Star pH with LogR Temperature Module Reference Guide...
Страница 6: ...This page intentionally left blank...
Страница 12: ...8 Thermo Scientific Orion Versa Star pH with LogR Temperature Module This page intentionally left blank...
Страница 18: ...14 Thermo Scientific Orion Versa Star pH with LogR Temperature Module This page intentionally left blank...
Страница 20: ...16 Thermo Scientific Orion Versa Star pH with LogR Temperature Module This page intentionally left blank...
Страница 38: ...34 Thermo Scientific Orion Versa Star pH with LogR Temperature Module Notes...
Страница 39: ...This page intentionally left blank...