Thermo Scientific Orion Star
™
and Star Plus Meter User Guide
EN-48
ISE Temperature Display and Calibration
ISE Temperature Display
Star Plus meters allow the temperature to be viewed on individual measurement
lines in addition to the temperature display on the top, left of the screen .
To view the temperature for the ISE measurement line:
1 . Press and hold
until the meter displays the measurement mode .
2 . Press
to choose the top display line . The arrow icon will point to the
selected line .
3 . Press
/
to change the value on the selected line . The top line
can be changed to display pH (pH), millivolts (mV), relative millivolts (RmV),
concentration (ISE), temperature (no icon) or a blank line .
ISE Temperature Calibration
The temperature calibration mode of the Star Plus meter allows the temperature
on each measurement line to be manually adjusted .
To calibrate the temperature for the ISE measurement line:
1 . In the measurement mode, press
to choose the top measurement line
and press
/
until the temperature is shown for the selected line .
2 . Press
to begin the calibration .
3 . When the reading stabilizes, the arrow icon and the first digit will flash .
Enter the temperature by pressing
/
to adjust each digit and
to
move to the next digit .
4 . Press
to save and end the calibration .
ISE Technique