73
Thermo Scientific Orion Star LogR™ Meter User Guide
Autosampler Interfacing
Pausing the Autosampler
To pause the autosampler during operation, press
. To resume operation, press
again.
Stopping the Autosampler
To stop the autosampler during operation and abort the run, press and hold
for
about 3 seconds.
Moving the Electrode Arm
At the end of a run, when the electrode arm has returned
to home position and is lowered into beaker one, press
to lift the electrode arm and press
again to
lower the electrode arm. See the diagram to the right. This
action is operational only when the autosampler is not
running and the electrode arm is in the home position.
Turning Off the Autosampler Function
To disable the autosampler, press
in the measurement mode to enter the setup
menu. Press
until
is displayed on top line. Press
until the arrow
icon points to the middle line and press
until
is displayed. Press
until
the arrow icon points to the bottom line and press
until
OFF
is displayed. Press
until the arrow icon points to the top line. Press
to return to measurement
mode.
Autosampler Dip Switch Settings
The dip switch cover is located on
the rear panel of the autosampler
and a small Phillips head screw
driver is needed to remove the dip
switch cover. The ON position for the
dip switches is down and the OFF
position is up. To set the autosampler
dip switches to 9600 for Com 1,
adjust DIP switch 4 and DIP switch 5
to the OFF position (up).
DIP 1
DIP 2
DIP 3
DIP 4
DIP 5
Even Parity
ON
No Parity
OFF
9600 Baud
OFF
OFF
1200 Baud
ON
OFF
2400 Baud
OFF
ON
38500
Baud
ON
ON
Содержание Orion Star LogR Meter
Страница 1: ...Orion Star LogR Meter User Guide...