Thermo Scientific Orion Star LogR™ Meter User Guide
Chapter IX
ISE Technique
The Orion Star LogR pH/ISE meter, Cat. No. 3115000, is capable of measuring
ISE values. When reading or calibrating ISE, the meter automatically switches the
temperature reading and compensation setting to
, or
if ATC probe is
not connected. (The temperature reading and compensation setting is shown on
the third line of the display.) Refer to the Selecting the Temperature Reading and
Compensation Setting section to change this setting if needed.
ISE Setup Menu
Note:
Refer to the Setup Menu section for the Setup Menu Table, which contains
a complete list of meter setup options and descriptions. Refer to Appendix A for a
description of the special setup menu features.
1. In the measurement mode, press
.
2. Press
/
to scroll through the setup menu until
is displayed on the
top line.
3. Press
to accept the selection and move the arrow icon to the middle line.
4. Press
/
to scroll through
for the ISE measurement resolution,
for the ISE measurement units,
for the ISE calibration range and
for the non-linear blank correction feature.
5. Press
to accept the selection and move the arrow icon to the bottom line.
6. To scroll through a list of options on the bottom line, press
/
until the
desired option is displayed. To enter a numeric value for an option on the bottom
line, press
/
to adjust each digit and
to move to the next digit.
7. Press
to accept the selection and move the arrow icon to the top line.
8. Repeat steps 3 through 7 to change another ISE setting or press
to return
to the measurement mode.