39
Thermo Scientific Orion Star LogR™ Meter User Guide
Data Archival and Retrieval
Datalog Deletion Setting
The datalog deletion setting determines if the meter will automatically delete
the datalog after it is downloaded to a printer or computer and if the meter will
overwrite the datalog points when the datalog is full. If the datalog deletion setting
is set to
, the meter will automatically delete the datalog after the datalog is
downloaded to a printer or computer. The meter will also display the
error message when all 1000 datalog points are filled and the datalog must be
downloaded to a printer or computer to clear the error message. If the datalog
deletion setting is set to
, the meter will overwrite the oldest datalog point when
all 1000 datalog points are filled and will not delete the datalog after the datalog is
downloaded to a printer or computer.
1. In the measurement mode, press
.
2. Press
/
until
is displayed on top line.
3. Press
to accept the selection and move the arrow icon to the middle line
and press
/
until
is displayed.
4. Press
to accept the selection and move the arrow icon to the bottom line
and press
/
until
or
is displayed.
5. Press
to accept the selection and move the arrow icon to the top line.
6. Press
to save the setup option and return to measurement mode.
Note:
If the datalog is not required, set the datalog deletion setting to
to
prevent the error 038 (datalog full) message.