Chapter 7
|
Data Transfer and Software Updates
Star A200-A300 Meter Remote Control Commands
Press the Esc key on the computer keypad to stop a command being executed.
Command
Command Details
GETMEAS
GETMEAS <CR>
Prints the current channel measurement immediately.
GETMEAS Data Count <CR>
Prints the current channel measurement for a set number of times.
Example: GETMEAS 2
GETMEASTIMED CH_Channel Combination, Time Interval <CR>
Set the channel(s) to be measured and the time interval for the measurement,
measurements are sent to the meter data log
Channel Combination: Channel numbers to be measured.
Time Interval: The measurement retrieving time interval in seconds.
Example: GETMEASTIMED CH_12 5 <CR>
STOP <CR>
Command exits Timed measurement.
GETCAL
GETCAL MODE <CR>
Prints all the channel mode calibration data. If no calibration is available, will return
“>” to receive next command.
Example: GETCAL MODE <CR>
GETCAL MODE <CR>
Print calibration data for specific MODE as:
PH
COND
DO
RMV
RES
RDO
ORP
SALT
ISE
TDS
Example: GETCAL PH <CR>
GETLOG
GETLOG <CR>
Prints all the logged measurement data. If no data is logged, will return “>” to
receive next command.
Example: GETLOG <CR>
GETLOG START, END <CR>
Prints the data in a specified range. If no data is available, will return “>” to receive
next command.
Example: GETLOG 10 100 <CR>
SYSTEM
SYSTEM <CR>
Prints the Meter Model, Serial Number, Software Version
Example: 329, 12345, 2.53
SETRTC
SETRTC YY MM DD HH MM SS <CR>
Set the date and time (in 24 hour format) for the meter
Example: SETRTC 13 07 15 08 30 00
68
|
Orion Star A220 Series Portable Meter User Manual
Thermo Scientific