Chapter 3
|
Meter Setup Menus
Mode and Settings Menu for DO/RDO Channel
Use the Mode and Settings menu to customize settings for the dissolved oxygen channel of
Orion Star A223 RDO/DO meters. Default meter settings are bold.
Parameter
Settings
Details
Measure Mode
•
DO – Polarographic
•
RDO
•
Auto
Set the type of dissolved oxygen sensor connected to
the meter as DO polarographic or RDO optical or
select Auto and the meter will automatically detect
which type of dissolved oxygen sensor is connected
and update the displayed measurement mode
Measure Unit
•
mg/L
•
% sat
Set the units displayed with the dissolved oxygen
measurement value
Resolution
(% saturation)
1 Decimal Place
2 Decimal Places
1
0.1
Set the resolution of the dissolved oxygen
measurement value as percent saturation
Resolution
(mg/L)
1 Decimal Place
2 Decimal Places
0.1
0.01
Set the resolution of the dissolved oxygen
measurement value as mg/L
Read Type
•
Auto
•
Continuous
Set the read type to determine how the meter takes a
measurement and when the measurement is sent to
the data log and/or external device
Baro Press
•
Auto
•
Manual 760.0mmHg
Set the automatic barometric pressure compensation
source as the internal barometer (Auto) or a manually
entered barometric pressure value (Manual)
Salinity Correct
•
Manual 0.0 Enter the salinity value of samples for automatic
salinity correction of dissolved oxygen measurements
Cal Due Alarm
•
Cal Due Alarm
On
Set the calibration due alarm off or on, when the
alarm is on, the fixed calibration time interval is 12
hours and the alarm is triggered after 12 hours has
elapsed without a calibration being performed
RDO Cap Life
(RDO only)
RDO Cap Life
Serial Number
0.0
0
View the remaining RDO optical DO sensor cap life in
days and the serial number
Barometric Pressure Compensation Setting
Orion Star A223 RDO/DO meters have an internal barometer that can be used for automatic
pressure compensated dissolved oxygen readings. Manual entry of the barometric pressure
can be used for automatic pressure compensated dissolved oxygen readings, which is useful
when dissolved oxygen is measured with a submerged sensor or in a pressurized vessel. The
pressure value must be entered as mm Hg.
1 mm Hg = 0.03937 inch Hg = 1.3332 hPa (mBar) = 0.01934 PSI
Thermo Scientific
Orion Star A220 Series Portable Meter User Manual
|
27