Chapter 5
|
Using the Conductivity Channel
Conductivity Standards vs. Temperature Table
Temperature
(°C)
Orion 111.9mS
Standard (mS/cm)
Orion 12.9mS
Standard (mS/cm)
Orion 1413μS
Standard (μS/cm)
Orion 147μS
Standard (μS/cm)
Orion 100μS
Standard (μS/cm)
0
65.10
7.135
776
81
54
1
66.84
7.344
799
83
56
2
68.59
7.555
822
86
58
3
70.35
7.768
846
88
59
4
72.12
7.983
870
91
61
5
73.91
8.200
894
93
63
6
75.70
8.418
918
96
64
7
77.50
8.638
943
98
66
8
79.32
8.860
968
101
68
9
81.15
9.084
992
103
70
10
82.98
9.309
1017
106
72
11
84.83
9.535
1043
108
73
12
86.69
9.763
1068
111
75
13
88.56
9.993
1094
114
77
14
90.45
10.22
1119
116
79
15
92.34
10.46
1145
119
81
16
94.24
10.69
1171
122
83
17
96.15
10.93
1198
125
85
18
98.08
11.16
1224
127
87
19
100.0
11.40
1251
130
88
20
102.0
11.64
1277
133
90
21
103.9
11.88
1304
136
92
22
105.9
12.12
1331
138
94
23
107.9
12.36
1358
141
96
24
109.9
12.61
1386
144
98
25
111.9
12.85
1413
147
100
26
113.9
13.10
1441
150
102
27
115.9
13.35
1468
153
104
28
117.9
13.59
1496
156
106
29
120.0
13.84
1524
159
108
30
122.0
14.09
1552
161
110
31
124.1
14.34
1580
164
112
32
126.2
14.59
1608
167
114
33
128.3
14.85
1636
170
117
34
130.4
15.10
1665
173
119
35
132.5
15.35
1693
176
121
36
134.6
15.61
1722
179
123
37
136.7
15.86
1751
182
125
38
138.9
16.12
1780
185
127
39
141.0
16.37
1808
188
129
40
143.2
16.63
1837
191
131
41
145.4
16.89
1866
194
134
42
147.6
17.15
1896
197
136
43
149.8
17.40
1925
200
138
44
152.0
17.66
1954
203
140
45
154.2
17.92
1983
206
142
46
156.4
18.18
2013
209
145
47
158.7
18.44
2042
212
147
48
160.9
18.70
2071
215
149
49
163.2
18.96
2101
219
151
50
165.4
19.22
2130
222
154
Thermo Scientific
Orion Star A220 Series Portable Meter User Manual
|
45
Содержание Orion Star A220 Series
Страница 5: ......