- 15 -
Calibration
Loop
To
enter
this
loop
you
must
be
in
the
operators
loop
and
displaying
the
temperature.
Depress
and
hold
the
ENTER
key.
While
holding
the
ENTER
key
enter
the
key
sequence
YES-NO-YES.
Use
this
loop
to
calibrate
the
RTD,
resistivity,
the
range
of
the
resistivity
out
signal
(V/Mohm),
the
range
of
the
temperature
out
signal
(100mV/°C)
and
the
high
and
low
flow
calibration
parameters.
Scroll
through
the
CALIBRATE
menu
using
the
NEXT
or
LAST
key,
press
ENTER
to
go
to
the
desired
sub-menu.
Scroll
through
the
sub-menu
options
using
the
NEXT
or
LAST
key,
change
the
values
using
the
YES
or
NO
key.
Leave
the
sub-menu
using
the
NEXT
or
LAST
key.
The
display
will
indicate
the
save
prompt,
press
YES
or
NO.
Figure
4
Calibration
Loop
3)
RESIST
OUT?